Assessment of spatial uncertainty in computed tomography-based Gamma Knife stereotactic radiosurgery process with automated positioning system
Authors:
- 16 July 2014
- journal article
- research article
- Published by Springer Nature in Acta Neurochirurgica
- Vol. 156 (10) , 1929-1935
Abstract
In this study, we assessed the geometric accuracy of an automated positioning system in Gamma Knife (GK) surgery. Specifically, we looked at the total spatial uncertainty over the entire treatment range of GK stereotactic radiosurgery (SRS) procedures in both the GK model C and the Perfexion (PFX). An originally-developed phantom and a radiochromic film were used for obtaining actual dose distributions. The phantom, with inserted films on different axial planes (z = 60, 75, 100, 125, 140 mm), sagittal planes (x = 60, 75, 100, 125, 140 mm), and coronal planes (y = 60, 75, 100, 125, 140 mm), was placed on a Leksell skull frame. Computed tomography (CT) was then performed with a stereotactic localizer box attached to the frame, and dose planning was made using the Leksell GammaPlan treatment planning system. The phantom finally received beam delivery using a single shot of a 4-mm collimator helmet. The discrepancy between the planned shot position and the irradiated center position was evaluated by a dedicated film analysis software. The total uncertainty of CT-based GK SRS was less than 1 mm for almost all measured points over the stereotactic space in both the model C and the PFX. In addition, the geometric accuracy of the automated positioning system was estimated to be less than 0.1 mm and equal to 0.5 mm in the central and peripheral areas, respectively. We confirmed that the total spatial uncertainties of both the GK model C and the PFX are acceptable for clinical use.Keywords
This publication has 16 references indexed in Scilit:
- Long-term quality of life and tumour control following gamma knife radiosurgery for vestibular schwannomaActa Neurochirurgica, 2013
- Assessment and characterization of the total geometric uncertainty in Gamma Knife radiosurgery using polymer gelsMedical Physics, 2013
- Assessment of Spatial Uncertainties in the Radiotherapy Process With the Novalis SystemInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- RADIOSURGERY WITH THE WORLD'S FIRST FULLY ROBOTIZED LEKSELL GAMMA KNIFE PERFEXION IN CLINICAL USENeurosurgery, 2009
- THE LEKSELL GAMMA KNIFE PERFEXION AND COMPARISONS WITH ITS PREDECESSORSNeurosurgery, 2008
- Clinical Evaluation of Targeting Accuracy of Gamma Knife Radiosurgery in Trigeminal NeuralgiaInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2007
- Stereotactic Imaging for Radiosurgery: Localization Accuracy of Magnetic Resonance Imaging and Positron Emission Tomography Compared with Computed TomographyStereotactic and Functional Neurosurgery, 2006
- Quality assurance in stereotactic space. A system test for verifying the accuracy of aim in radiosurgeryMedical Physics, 2002
- Quality assurance for the Leksell gamma unit: Considering magnetic resonance image-distortion and delineation failure in the targeting of the internal auditory canalMedical Physics, 1999
- Physics of gamma knife approach on convergent beams in stereotactic radiosurgeryInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1990