LiF:Mg,Ti TLD response as a function of photon energy for moderately filtered x‐ray spectra in the range of 20–250 kVp relative to
Authors:
- A. A. NunnANA. A. Nunn
- S. D. DavisSDS. D. Davis
- J. A. MickaJMJ. A. Micka
- L. A. DeWerdLDL. A. DeWerd
- 17 April 2008
- journal article
- Published by Wiley in Medical Physics
- Vol. 35 (5) , 1859-1869
Abstract
The response of LiF:Mg,Ti thermoluminescent dosimeters (TLDs) as a function of photon energy was determined using irradiations with moderately filtered x‐ray beams in the energy range of 20–250 kVp relative to the response to irradiations with photons. To determine if the relative light output from LiF:Mg,Ti TLDs per unit air kerma as a function of photon energy can be predicted using calculations such as Monte Carlo (MC) simulations, measurements from the x‐ray beam irradiations were compared with MC calculated results, similar to the methodology used by Davis et al. [Radiat. Prot. Dosim. 106, 33–43 (2003)]. TLDs were irradiated in photon beams with well‐known air kerma rates using the National Institute of Standards and Technology traceable M‐series x‐ray beams in the range of 20–250 kVp. For each x‐ray beam, several sets of TLDs were irradiated for times corresponding to different air kerma levels to take into account any dose nonlinearity. TLD light output was then compared to that from several sets of TLDs irradiated at similar corresponding air kerma levels using a irradiator. The MC code MCNP5 was used to account for photon scatter and attenuation in the holder and TLDs and was used to calculate the predicted relative TLD light output per unit air kerma for irradiations with each of the experimentally used photon beams. The measured relative TLD response as a function of photon energy differed by up to from the MC calculations. We conclude that MC calculations do not accurately predict the relative response of TLDs as a function of photon energy, consistent with the conclusions of Davis et al. [Radiat. Prot. Dosim. 106, 33–43 (2003)]. This is likely due to complications in the solid state physics of the thermoluminescence process that are not incorporated into the simulation.Keywords
This publication has 33 references indexed in Scilit:
- Supplement to the 2004 update of the AAPM Task Group No. 43 ReportMedical Physics, 2007
- The effects of ionisation density on the thermoluminescence response (efficiency) of LiF:Mg,Ti and LiF:Mg,Cu,PRadiation Protection Dosimetry, 2004
- Changes in the US primary standards for the air kerma from gamma-ray beamsJournal of Research of the National Institute of Standards and Technology, 2003
- Theory of thermoluminescence gamma dose response: The unified interaction modelNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 2001
- Thermoluminescent dosimetry of the Symmetra™ 125I model I25.S06 interstitial brachytherapy seedMedical Physics, 2001
- Photon Energy Response Matrix for Environmental Monitoring Systems Based on LiF:Mg,Ti and Hypersensitive Phosphors (LiF:Mg,Cu,P and a-Al2O3:C)Radiation Protection Dosimetry, 1999
- Influence of the variation of energy spectra with depth in the dosimetry of192Ir using LiF TLDPhysics in Medicine & Biology, 1988
- The three-dimensional dose distribution of125I seeds in tissuePhysics in Medicine & Biology, 1983
- Determination of W Values and Backscatter Coefficients for Slow Electrons in AirRadiation Research, 1978
- Thermoluminescent Radiation Dosimetry Utilizing LiFHealth Physics, 1964