Discerning aortic waves during intra-aortic balloon pumping and their relation to benefits of counterpulsation in humans
Open Access
- 1 November 2009
- journal article
- research article
- Published by American Physiological Society in Journal of Applied Physiology
- Vol. 107 (5) , 1497-1503
Abstract
An explanation of the mechanisms leading to the beneficial hemodynamic effects of the intra-aortic balloon pump (IABP) is lacking. We hypothesized that inflation and deflation of the balloon would generate a compression (BCW) and an expansion (BEW) wave, respectively, which, when analyzed with wave intensity analysis, could be used to explain the hemodynamic benefits of IABP support. Simultaneous ascending aortic pressure (Pao) and flow rate (Qao) were recorded in 25 patients during control conditions and with IABP support of 1:1 and 1:2. Diastolic aortic pressure augmentation (Paug) and end-diastolic aortic pressure (ED Pao) reduction were calculated from Pao. Energies of the BCW and BEW were obtained by integrating the wave intensity contour over time. Paug was 19.1 mmHg (SD 13.6) during 1:2 support. During 1:1 support significantly higher Paug of 21.1 mmHg (SD 13.4) was achieved ( P < 0.001). ED Pao decreased from 50.9 mmHg (SD 15.1) to 43.9 mmHg (SD 15.7) ( P < 0.0001) during 1:1 assistance and the decrease was not statistically different with 1:2. During 1:1 support the energy of BCW was correlated positively to Paug ( r = 0.83, P < 0.0001) and energy of the BEW correlated negatively to ED Pao ( r = 0.78, P < 0.005); these relationships were not statistically different during 1:2. In conclusion, the energies of the BCW and BEW are directly related to Paug and ED Pao, which are the conventional hemodynamic parameters indicating IABP benefits. These findings imply a cause and effect mechanism between the energies of BCW and BEW, and IABP hemodynamic effects.Keywords
This publication has 28 references indexed in Scilit:
- Aortic wave intensity analysis of ventricular-vascular interaction during incremental dobutamine infusion in adult sheepAmerican Journal of Physiology-Heart and Circulatory Physiology, 2008
- Effects of intra-aortic balloon counterpulsation on coronary pressure in patients with stenotic coronary arteriesAmerican Heart Journal, 2007
- Pressure and flow across severe stenoses: Does the intra-aortic balloon pump do its job?American Heart Journal, 2007
- Evidence of a Dominant Backward-Propagating “Suction” Wave Responsible for Diastolic Coronary Filling in Humans, Attenuated in Left Ventricular HypertrophyCirculation, 2006
- Wave-energy patterns in carotid, brachial, and radial arteries: a noninvasive approach using wave-intensity analysisAmerican Journal of Physiology-Heart and Circulatory Physiology, 2005
- Beat-to-Beat Effects of Intraaortic Balloon Pump Timing on Left Ventricular Performance in Patients With Low Ejection FractionThe Annals of Thoracic Surgery, 2005
- Impact of intra-aortic balloon counterpulsation with different balloon volumes on cardiac performance in humansCatheterization and Cardiovascular Interventions, 2002
- Determination of wave speed and wave separation in the arteriesJournal of Biomechanics, 2001
- Treating severe cardiogenic shock by large counterpulsation volumesThe Annals of Thoracic Surgery, 1996
- Forward and Backward Running Waves in the Arteries: Analysis Using the Method of CharacteristicsJournal of Biomechanical Engineering, 1990