Fatigue Analysis of Plates Used for Fracture Stabilization in Small Dogs and Cats
Authors:
- 3 August 2006
- journal article
- research article
- Published by Wiley in Veterinary Surgery
- Vol. 35 (6) , 573-578
Abstract
To evaluate the fatigue life of stacked and single, veterinary cuttable plates (VCP) and small, limited contact, dynamic compression plates (LC-DCP). In vitro biomechanical study. Fracture models (constructs; n = 8) were assembled for each of 6 groups all with 8-hole plates: 2.0 mm LC-DCP; 2.4 mm LC-DCP; single 1.5/2.0 mm VCP; stacked 1.5/2.0 mm VCP; single 2.0/2.7 mm VCP; and stacked 2.0/2.7 mm VCP. Plate(s) were secured to 2 polyvinylchloride pipe lengths, mounted in a testing system with a custom jig, and subjected to axial loading (10-100 N) for 1,000,000 cycles at 10 Hz or until failure. Differences in number of cycles to failure among groups were compared. Failure mode was determined. All LC-DCP and single VCP constructs failed before 1,000,000 cycles. Stacked 2.0/2.7 mm VCP constructs withstood 1,000,000 cycles without failure. ANOVA and Fisher's least significant difference tests demonstrated significantly more cycles to failure for the stacked 1.5/2.0 mm VCP and stacked 2.0/2.7 mm VCP compared with the single 1.5/2.0 mm VCP, single 2.0/2.7 mm VCP, 2.0 mm LC-DCP, or 2.4 mm LC-DCP. Constructs that failed did so through a screw hole adjacent to the gap. Stacked VCP constructs have greater fatigue lives than comparably sized LC-DCP or single VCP constructs. Plates with 2.4 mm screws were not significantly different from the comparable construct with 2.0 mm screws. Although these data reveal that stacked VCP create a superior construct with respect to cyclic fatigue, surgeons must decide whether this is a clinical advantage on a case-by-case basis.Keywords
This publication has 8 references indexed in Scilit:
- Bending properties of stainless steel dynamic compression plates and limited contact dynamic compression platesVeterinary and Comparative Orthopaedics and Traumatology, 2001
- Manual of Internal Fixation in Small Animals1998
- A Mechanical Comparison of the Dynamic Compression Plate, Limited Contact-Dynamic Compression Plate, and Point Contact FixatorJournal of Orthopaedic Trauma, 1995
- The limited contact dynamic compression plate (LC-DCP)Archives of orthopaedic and trauma surgery, 1990
- Internal Fixation with a Veterinary Cuttable Plate in Small AnimalsVeterinary and Comparative Orthopaedics and Traumatology, 1989
- The Effects of Screw Removal on Bone Strain in an Idealized Plated Bone ModelVeterinary Surgery, 1988
- Fatigue fracture of bone platesInjury, 1980
- Fatigue behavior of type 316 stainless steel under simulated body conditionsJournal of Biomedical Materials Research, 1971