Abstract
Background: Repeat patient testing‐quality control (RPT‐QC) uses retained patient samples as an alternative to commercial quality control material (QCM). We elected to calculate and validate RPT‐QC limits for red blood cell count (RBC), hemoglobin (HBG), hematocrit (HCT), and white blood cell count (WBC).Objectives: (1) To validate RPT‐QC across a network of four harmonized Sysmex XT‐2000iV hematology analyzers and determine the total error that can be controlled with RPT‐QC. (2) To generate quality control (QC) limits using the standard deviation (SD) of the duplicate measurement differences and determine a suitable simple QC rule with a probability of error detection >0.85 and probability of false rejection 5.5 for RBC, HGB, and WBC, but not for HCT.