Feasibility of equivalent performance of 3D TOF [18F]-FDG PET/CT with reduced acquisition time using clinical and semiquantitative parameters
Open Access
- 1 May 2021
- journal article
- research article
- Published by Springer Nature in EJNMMI Research
- Vol. 11 (1) , 1-11
Abstract
Background: High-performance time-of-flight (TOF) positron emission tomography (PET) systems have the capability for rapid data acquisition while preserving diagnostic image quality. However, determining a reliable and clinically applicable cut-off of the acquisition time plays an important role in routine practice. This study aimed to assess the diagnostic equivalence of short acquisition time of 57 with routine 75 seconds per bed position (s/BP) of [18F]-fluoro-deoxy-glucose (FDG) PET. Phantom studies applying EARL criteria suggested the feasibility of shortened acquisition time in routine clinical imaging by 3D TOF PET/CT scanners. Ninety-six patients with melanoma, lung or head and neck cancer underwent a standard whole-body, skull base-to-thigh or vertex-to-thigh [18F]-FDG PET/CT examination using the 3D TOF Ingenuity TF PET/CT system (Philips, Cleveland, OH). The [18F]-FDG activity applied was equal to 4MBq per kg body weight. Retrospectively, PET list-mode data were used to calculate a second PET study per patient with a reduced acquisition time of 57 s instead of routine 75 s/BP. PET/CT data were reconstructed using a 3D OSEM TOF algorithm. Blinded patient data were analysed by two nuclear medicine physicians. The number of [18F]-FDG-avid lesions per body region (head&neck, thorax, abdomen, bone, extremity) and image quality (grade 1–5) were evaluated. Semiquantitative analyses were performed by standardized uptake value (SUV) measurements using 3D volume of interests (VOI). The visual and semiquantitative diagnostic equivalence of 214 [18F]-FDG-avid lesions were analysed in the routine standard (75 s/BP) as well as the calculated PET/CT studies with short acquisition time. Statistical analyses were performed by equivalence testing and Bland–Altman plots. Results: Lesion detection rate per patient’s body region agreed in > 98% comparing 57 s/BP and 75 s/BP datasets. Overall image quality was determined as equal or superior to 75 s in 80% and 69%, respectively. In the semiquantitative lesion-based analyses, a significant equivalence was found between the 75 s/BP and 57 s/BP PET/CT images both for SUVmax (p = 0.004) and SUVmean (p = 0.003). Conclusion: The results of this study demonstrate significant clinical and semiquantitative equivalence between short acquisition time of 57 s/BP and standard 75 s/BP 3D TOF [18F]-FDG PET/CT scanning, which may improve the patient’s workflow in routine practice.Keywords
Funding Information
- Paracelsus Medizinische Privatuniversität (D-17/02/007-PIL)
This publication has 30 references indexed in Scilit:
- FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0European Journal of Nuclear Medicine and Molecular Imaging, 2014
- Japanese guideline for the oncology FDG-PET/CT data acquisition protocol: synopsis of Version 2.0Annals of Nuclear Medicine, 2014
- Evaluation of noise equivalent count parameters as indicators of adult whole-body FDG-PET image qualityAnnals of Nuclear Medicine, 2013
- A retrospective evaluation of radiation dose associated with low dose FDG protocols in whole-body PET/CTAustralasian Physics & Engineering Sciences in Medicine, 2011
- Time-of-flight PET/CT using low-activity protocols: potential implications for cancer therapy monitoringEuropean Journal of Nuclear Medicine and Molecular Imaging, 2010
- FDG PET and PET/CT: EANM procedure guidelines for tumour PET imaging: version 1.0European Journal of Nuclear Medicine and Molecular Imaging, 2009
- Impact of Time-of-Flight on PET Tumor DetectionJournal of Nuclear Medicine, 2009
- Comparison of Imaging Protocols for 18F-FDG PET/CT in Overweight Patients: Optimizing Scan Duration Versus Administered DoseJournal of Nuclear Medicine, 2009
- Standards for PET Image Acquisition and Quantitative Data AnalysisJournal of Nuclear Medicine, 2009
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986