Abstract
Our study shows that the size and shape of coagulation necrosis created by the bipolar radiofrequency ablation device tested in this study is not affected significantly by changes in portal vein blood flow rates. PurposeTo investigate the effect of change in portal vein flow rates on the size and shape of ablations created by a bipolar radiofrequency (RF) ablation device.Materials and MethodsThis study was exempt from institutional animal care and use committee review. An in vitro bovine liver model perfused with autologous blood via the portal vein at three flow rates (60, 80, 100 mL/min per 100 g of liver) was used. Four ablations, two bipolar and two monopolar (control probe), were made in each of five livers perfused at each flow rate. Short- and long-axis diameters were measured from gross specimens, and volume and sphericity index were calculated for each ablation. A general linear mixed model accounting for correlation within the liver was used to evaluate the effects of flow on ablations. Analyses were performed by using software.ResultsThere was no significant difference in the size or shape of ablations created by the bipolar device at the different flow rates (P > .05 for all outcomes). The monopolar de...

This publication has 41 references indexed in Scilit: