Target Indices for Clinical Ablation in Atrial Fibrillation
Authors:
- 1 February 2014
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation: Arrhythmia and Electrophysiology
- Vol. 7 (1) , 63-68
Abstract
Background—: In animal studies of radiofrequency ablation, lesion sizes plateau as the maximum lesion size is reached for an ablation. Lesion parameters are not available in clinical ablations, but preclinical work suggests that these correlate with impedance drop and electrogram attenuation. Characterization of the relationships between catheter contact force, ablation duration, and these surrogate markers of lesion formation may allow us to define targets for effective ablation. Methods and Results—: Fifteen patients undergoing first-time radiofrequency ablation for nonparoxysmal atrial fibrillation were studied. All were in atrial fibrillation at the time of the procedure. Ablations were performed with an irrigated-tip contact force–sensing catheter in temperature-controlled mode (temperature limited to 48°C, power to 30 W). Included were 285 left atrial static ablations, 247 with additional impedance data. The ablation force time integral (FTI) correlated with the attenuation of the electrogram with ablation (Spearman ρ, –0.14; P =0.02): the relationship plateauing from 500 g·s, a reduction in the electrogram amplitude of 20%. The FTI also correlated with the impedance drop during ablation (Spearman ρ, 0.79; P <0.0005): the relationship was logarithmic, the reduction in the impedance with an increasing FTI also plateauing from 500 g·s, an impedance drop of 7.5%. The ablation duration affected the impedance drop at an FTI if the duration was <10 s. Beyond this time point, the FTI achieved rather than the ablation duration or mean contact force applied determined the impedance drop. Conclusions—: During nonparoxysmal atrial fibrillation ablation, an FTI of 500 g·s should be targeted with ablation duration of ≥10 s. Clinical Trials Registration—: URL: http://clinicaltrials.gov/ . Unique Identifier: NCT01587404.Keywords
This publication has 20 references indexed in Scilit:
- Contact force sensing technology identifies sites of inadequate contact and reduces acute pulmonary vein reconnection: A prospective case control studyInternational Journal of Cardiology, 2013
- Ablation Effect Indicated by Impedance Fall is Correlated with Contact Force Level During Ablation for Atrial FibrillationJournal of Cardiovascular Electrophysiology, 2013
- Predictive value of impedance changes for real-time contact force measurements during catheter ablation of atrial arrhythmias in humansHeart Rhythm, 2013
- Electrical Reconnection After Pulmonary Vein Isolation Is Contingent on Contact Force During Initial TreatmentCirculation: Arrhythmia and Electrophysiology, 2013
- Robotic assistance and general anaesthesia improve catheter stability and increase signal attenuation during atrial fibrillation ablationEP Europace, 2012
- Improved Electrogram Attenuation during Ablation of Paroxysmal Atrial Fibrillation with the Hansen Robotic SystemPacing and Clinical Electrophysiology, 2012
- Local unipolar and bipolar electrogram criteria for evaluating the transmurality of atrial ablation lesions at different catheter orientations relative to the endocardial surfaceHeart Rhythm, 2010
- Effect of Electrode Orientation on Lesion Sizes Produced by Irrigated Radiofrequency Ablation CathetersJournal of Cardiovascular Electrophysiology, 2009
- Different Patterns of the Fall of Impedance as the Result of Heating During Ostial Pulmonary Vein Ablation: Implications for Power TitrationPacing and Clinical Electrophysiology, 2005
- Determinants of Lesion Size During Radiofrequency Catheter Ablation: The Role of Electrode‐Tissue Contact Pressure and Duration of Energy DeliveryJournal of Cardiovascular Electrophysiology, 1991