Quality of tri-Co-60 MR-IGRT treatment plans in comparison with VMAT treatment plans for spine SABR
- 8 December 2016
- journal article
- research article
- Published by Oxford University Press (OUP) in British Journal of Radiology
- Vol. 90 (1070) , 20160652
Abstract
Objective: To investigate the plan quality of tri-Co-60 intensity-modulated radiation therapy (IMRT) plans for spine stereotactic ablative radiotherapy (SABR). Methods: A total of 20 patients with spine metastasis were retrospectively selected. For each patient, a tri-Co-60 IMRT plan and a volumetric-modulated arc therapy (VMAT) plan were generated. The spinal cords were defined based on MR images for the tri-Co-60 IMRT, while isotropic 1-mm margins were added to the spinal cords for the VMAT plans. The VMAT plans were generated with 10-MV flattening filter-free photon beams of TrueBeam STx™ (Varian Medical Systems, Palo Alto, CA), while the tri-Co-60 IMRT plans were generated with the ViewRay™ system (ViewRay inc., Cleveland, OH). The initial prescription dose was 18 Gy (1 fraction). If the tolerance dose of the spinal cord was not met, the prescription dose was reduced until the spinal cord tolerance dose was satisfied. Results: The mean dose to the target volumes, conformity index and homogeneity index of the VMAT and tri-Co-60 IMRT were 17.8 ± 0.8 vs 13.7 ± 3.9 Gy, 0.85 ± 0.20 vs 1.58 ± 1.29 and 0.09 ± 0.04 vs 0.24 ± 0.19, respectively. The integral doses and beam-on times were 16,570 ± 1768 vs 22,087 ± 2.986 Gy cm3 and 3.95 ± 1.13 vs 48.82 ± 10.44 min, respectively. Conclusion: The tri-Co-60 IMRT seems inappropriate for spine SABR compared with VMAT. Advances in knowledge: For spine SABR, the tri-Co-60 IMRT is inappropriate owing to the large penumbra, large leaf width and low dose rate of the ViewRay system.Keywords
This publication has 26 references indexed in Scilit:
- RTOG 0631 phase 2/3 study of image guided stereotactic radiosurgery for localized (1-3) spine metastases: Phase 2 resultsPractical Radiation Oncology, 2014
- Stereotactic body radiotherapy for solitary spine metastasisRadiation Oncology Journal, 2013
- RapidArc®vsintensity-modulated radiation therapy for hepatocellular carcinoma: a comparative planning studyBritish Journal of Radiology, 2012
- Photon energy-modulated radiotherapy: Monte Carlo simulation and treatment planning studyMedical Physics, 2012
- Stereotactic body radiation therapy: The report of AAPM Task Group 101Medical Physics, 2010
- Clinical Applications of Volumetric Modulated Arc TherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2010
- Volumetric Modulated Arc Therapy: Planning and Evaluation for Prostate Cancer CasesInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2010
- Volumetric Arc Intensity–Modulated Therapy for Spine Body Radiotherapy: Comparison With Static Intensity-Modulated TreatmentInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119Medical Physics, 2009
- Radiation therapy oncology group: Radiosurgery quality assurance guidelinesInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1993