Determining the filtration efficiency of half-face medical protection mask (N99) against viral aerosol
Authors:
- 13 April 2010
- journal article
- Published by Springer Nature in Aerobiologia
- Vol. 26 (3) , 245-251
Abstract
Hospital-based outbreaks of severe acute respiratory syndrome (SARS) have once again highlighted the vulnerability of healthcare workers (HCWs). Use of personal respiratory protective equipment was the main method used by HCWs to avoid nosocomial transmission. This paper describes the technology used to evaluate the filtration efficiency of the half-face medical protection mask (N99), manufactured by Firmshield Biotechnology, against viral aerosol. Viral aerosol was generated and then sampled simultaneously with and without the test mask. This enables a percentage efficiency value to be calculated against test phage f2 aerosols (surrogates of viral pathogen aerosols). At the same time the mask filtration efficiency against NaCl particle aerosol was determined by use of TSI8130 equipment and face-fit factor was tested by use of TSI8020 equipment. The half-face medical protection mask (N99) evaluated by use of the viral aerosol had a filtration efficiency >99%. The mask filtration efficiency against NaCl particle aerosol was 99.634 ± 0.024% and it had a good face-fit factor. This half-face medical protection mask (N99) can protect the wearer from viral aerosol disease transmission. The test method can be used to assess filtration efficacy against viral aerosol of masks used for respiratory protection.Keywords
This publication has 21 references indexed in Scilit:
- Factors involved in the aerosol transmission of infection and control of ventilation in healthcare premisesJournal of Hospital Infection, 2006
- Determining the performance of a commercial air purification system for reducing airborne contamination using model micro-organisms: a new test methodologyJournal of Hospital Infection, 2005
- Protecting health care workers from SARS and other respiratory pathogens: A review of the infection control literatureAmerican Journal of Infection Control, 2005
- Respiratory protection against bioaerosols: Literature review and research needsAmerican Journal of Infection Control, 2004
- Selecting Respirators for Control of Worker Exposure to Infectious AerosolsInfection Control & Hospital Epidemiology, 1999
- Performance of N95 Respirators: Filtration Efficiency for Airborne Microbial and Inert ParticlesAIHA Journal, 1998
- Collection of three bacterial aerosols by respirator and surgical mask filters under varying conditions of flow and relative humidityAnnals of Occupational Hygiene, 1997
- Mycobacterial Aerosol Collection Efficiency of Respirator and Surgical Mask Filters under Varying Conditions of Flow and HumidityApplied Occupational and Environmental Hygiene, 1997
- Evaluation of single-use masks and respirators for protection of health care workers against mycobacterial aerosolsAmerican Journal of Infection Control, 1994
- An evaluation of the microbial retention performance of three ventilator-circuit filtersIntensive Care Medicine, 1994