An Autonomous Robotic Phlebotomy Device: Extended Analytical Validation and Real-World Hemolysis Assessment
Authors:
- Jeroen A. RoestJRJeroen A. Roest
- Luuk F. P. GiesenLGLuuk F. P. Giesen
- Thijs C. van HoltenTHThijs C. van Holten
- Martine M. L. DeckersMDMartine M. L. Deckers
- Helena ChonHCHelena Chon
- Frederique M. A. KoopmanFKFrederique M. A. Koopman
- Michael A. FourauxMFMichael A. Fouraux
- Judith M. E. P. GillisJGJudith M. E. P. Gillis
- Robert de JongeRJRobert de Jonge
Open Access
- 17 November 2026
- journal article
- research article
- Published by MDPI AG in Diagnostics
- Vol. 16 (22) , 3822
Abstract
Background/Objectives: Automated phlebotomy may improve preanalytical quality and access to venous blood testing. Prior work evaluating the Aletta Autonomous Robotic Phlebotomy Device (ARPD) demonstrated analytical equivalence. This study expanded the initial analytical validation to additional blood collection tubes and analytes and also assessed real-world hemolysis rates. Methods: The study was conducted in three hospital outpatient phlebotomy departments, including a site with a high prevalence of (hemato-)oncology patients. The within-subject analytical validation study included Greiner VACUETTE (Cohort 1) and BD Vacutainer (Cohort 2) serum, lithium heparin, and EDTA tubes. Equivalence of potassium, lactate dehydrogenase, hemolysis index, and complete blood count (CBC) with differential between paired ARPD-collected and manually collected samples was assessed using Passing–Bablok regression. Real-world ARPD hemolysis (>0.5 g/L free hemoglobin) rates were determined in a separate cohort (Cohort 3) and compared with site-specific manual hemolysis rates. Results: In the analytical validation study, 119 patients were included in the analysis (Cohort 1, n = 52; Cohort 2, n = 67). All tests met the predefined acceptability criteria. The hemolysis index was statistically significantly lower in ARPD-collected lithium heparin samples in Cohort 1. In real-world use (1368 ARPD venipunctures), the ARPD hemolysis rate was 0.0% (95% CI: 0.0–0.3%), significantly lower than manual phlebotomy (1.6%, p = 0.025). Conclusions: The study demonstrated analytical equivalence in a clinically complex population and found reduced ARPD hemolysis rates. The results further substantiate that the ARPD could improve patient care by reducing preanalytical errors.Keywords
Funding Information
- Vitestro
This publication has 15 references indexed in Scilit:
- Performance, Safety, and Patient Experience of an Autonomous Robotic Phlebotomy Device: A Multicenter TrialClinical Chemistry, 2026
- Recommendations from the IFCC Working Group on Laboratory Errors and Patient Safety for the Global Adoption of an Essential Quality Indicators Panel in Laboratory MedicineClinical Chemistry and Laboratory Medicine (CCLM), 2025
- The American Society for Clinical Pathology 2024 Vacancy Survey of medical laboratories in the United StatesAmerican Journal of Clinical Pathology, 2025
- Comparison of a Novel Thin-Walled 25-gauge Needle Push Button Blood Collection Set to a Standard 23-gauge Needle in a Cancer Patient PopulationThe Journal of Applied Laboratory Medicine, 2023
- Quality indicators in laboratory medicine: state-of-the-art, quality specifications and future strategiesClinical Chemistry and Laboratory Medicine (CCLM), 2022
- Causes, consequences and management of sample hemolysis in the clinical laboratoryClinical Biochemistry, 2017
- Phlebotomy, a bridge between laboratory and patientBiochemia Medica, 2016
- An Integrated View of Potassium HomeostasisNew England Journal of Medicine, 2015
- Incomplete filling of lithium heparin tubes affects the activity of creatine kinase and γ-glutamyltransferaseBritish Journal of Biomedical Science, 2012
- Evaluation of Different Mixing Procedures for K2 EDTA Primary Samples on Hematological TestingLaboratory Medicine, 2007