Simulation of atomic force microscopy operation via three-dimensional finite element modelling
Authors:
- D T GethinDGD T Gethin
- 15 January 2009
- journal article
- Published by IOP Publishing in Nanotechnology
- Vol. 20 (6) , 065702
Abstract
Numerical modelling of atomic force microscopy cantilever designs and experiments is presented with the aim of exploring friction mechanisms at the microscale. As a starting point for this work, comparisons between finite element (FE) models and previously reported mathematical models for stiffness calibration of cantilevers (beam and V-shaped) are presented and discrepancies highlighted. A colloid probe (comprising a plain cantilever on which a particle is adhered) model was developed, and its normal and shear interaction were investigated, exploring the response of the probe accounting for inevitable imperfections in its manufacture. The material properties of the cantilever had significant impact on both the normal response and the lateral response. The sensitivity of the mechanical response in both directions was explored and it was found to be higher in terms of normal rather than lateral sensitivity. In lateral measurements, generic response stages were identified, comprising a first stage of twisting, followed by lateral bending, and then slipping. This was present in the two cantilever types explored (beam and V-shaped). Additionally, a model was designed to explore the dynamic sensitivity by comparing the simulation of a hysteresis loop with a previously reported experiment, and the results show good agreement in the response pattern. The ability to simulate the scan over an inclined surface representing the flank of an asperity was also demonstrated.Keywords
This publication has 11 references indexed in Scilit:
- A complete analysis of the laser beam deflection systems used in cantilever-based systemsUltramicroscopy, 2007
- Simulation of dynamic modes of atomic force microscopy using a 3D finite element modelUltramicroscopy, 2006
- Finite element analysis of V‐shaped cantilevers for atomic force microscopy under normal and lateral force loadsSurface and Interface Analysis, 2006
- The determination of atomic force microscope cantilever spring constants via dimensional methods for nanomechanical analysisNanotechnology, 2005
- Normal and torsional spring constants of atomic force microscope cantileversReview of Scientific Instruments, 2004
- Young’s modulus of silicon nitride used in scanning force microscope cantileversJournal of Applied Physics, 2004
- Atomic force microscope studies of stainless steel: Surface morphology and colloidal particle adhesionJournal of Materials Science, 2001
- Chemical Force Microscopy: Exploiting Chemically-Modified Tips To Quantify Adhesion, Friction, and Functional Group Distributions in Molecular AssembliesJournal of the American Chemical Society, 1995
- Lateral, normal, and longitudinal spring constants of atomic force microscopy cantileversReview of Scientific Instruments, 1994
- Lateral Force Microscopy Study on the Shear Properties of Self-Assembled Monolayers of Dialkylammonium Surfactant on MicaLangmuir, 1994