A single lateral hinge screw increased resistance to varus stress after medial closing wedge distal femoral osteotomy in a Sawbones model: A biomechanical analysis
Open Access
- 1 October 2025
- journal article
- Published by Wiley in Journal of Experimental Orthopaedics
- Vol. 12 (4)
Abstract
Purpose: To determine the biomechanical effects of a lateral hinge screw on postoperative fracture after medial closing wedge distal femoral osteotomy (MCWDFO). It was hypothesized that adding a lateral hinge screw would significantly increase the failure strength of the hinge. Methods: Twelve Sawbone femurs were used for the study. A biplanar cut was made, a protective lateral hinge pin was placed, and a 5 mm‐closing wedge osteotomy was performed in each specimen. After osteotomy, the hinge pin was removed in six specimens, and in the remaining six specimens, the hinge pin was replaced with a lateral hinge screw. A varus load was then applied utilizing a load frame in a single load‐to‐failure test. Failure was defined as the point just before a substantial drop in the force versus displacement curve. Nonparametric Wilcoxon rank‐sum test was performed to compare maximum load and stiffness between the hinge screw and no hinge screw groups with p < 0.05 considered the threshold of statistical significance. Results: The maximum load to failure was significantly higher in the hinge screw group versus the control group, 440 ± 70 N versus 256 ± 107 N (mean ± standard deviation), respectively. Load to failure was 72% higher in the hinge screw group versus the no hinge screw group. No significant difference in stiffness was found between the groups. Conclusion: Using a Sawbone model, the current data showed that placement of a lateral hinge screw significantly increased resistance to varus stress following MCWDFO compared with a construct having no screw. Additionally, no significant difference in stiffness was observed between the group with the hinge screw and the group without the screw. Further investigation should assess the clinical benefit of a hinge screw in reducing surgical morbidity during the postoperative period for patients undergoing a distal femoral osteotomy. Level of Evidence: N/A.Keywords
This publication has 26 references indexed in Scilit:
- Large opening gaps, unstable hinge fractures, and osteotomy line below the safe zone cause delayed bone healing after open-wedge high tibial osteotomyKnee Surgery, Sports Traumatology, Arthroscopy, 2018
- Filling Open Screw Holes in the Area of Metaphyseal Comminution Does Not Affect Fatigue Life of the Synthes Variable Angle Distal Femoral Locking Plate in the AO/OTA 33-A3 Fracture Model.2018
- Surgical Treatment With Closing-Wedge Distal Femoral Osteotomy for Recurrent Patellar Dislocation With Genu ValgumThe American Journal of Sports Medicine, 2018
- Distal Femoral Osteotomy for the Valgus Knee: Medial Closing Wedge Versus Lateral Opening Wedge: A Systematic ReviewArthroscopy, 2016
- Opening- and Closing-Wedge Distal Femoral OsteotomyOrthopaedic Journal of Sports Medicine, 2016
- Clinical and radiologic outcomes associated with the use of dynamic locking screws (DLS) in distal tibia fracturesEuropean Journal of Trauma and Emergency Surgery, 2015
- Varization open-wedge osteotomy of the distal femur: comparison between locking plate and angle blade plate constructsKnee Surgery, Sports Traumatology, Arthroscopy, 2014
- Wedge volume and osteotomy surface depend on surgical technique for distal femoral osteotomyKnee Surgery, Sports Traumatology, Arthroscopy, 2012
- Fractures Around the Lateral Cortical Hinge After a Medial Opening‐Wedge High Tibial Osteotomy: A New Classification of Lateral Hinge FractureArthroscopy, 2011
- Effects of Construct Stiffness on Healing of Fractures Stabilized with Locking PlatesJournal of Bone and Joint Surgery, 2010