Software-based simulation for pipeline vantage flow diverter preprocedural assessment: Method and validation study
- 23 July 2025
- journal article
- Published by SAGE Publications in Interventional Neuroradiology
Abstract
Background: Flow diverters (FDs) have revolutionized intracranial aneurysm (IA) treatment. Proper FD sizing is crucial for aneurysm occlusion and complication prevention. Ankyras (Mentice, Gothenburg, Sweden) is a device-specific sizing software. We evaluated Ankyras’ performance in predicting the final deployment of the Pipeline™ Vantage Embolization Device with Shield Technology™ (PVST) (Medtronic Neurovascular, Irvine, California, USA). Methods: We analyzed with Ankyras software the three-dimensional rotational angiography (3DRA) images of 10 consecutive patients with unruptured IA treated with PVST. Conventional digital subtraction angiography (DSA) measurements were used for FD sizing, FD was implanted, and postprocedure DSA and 3DRA images were obtained. Ankyras software generated an aneurysm model and simulated the FD size used in actual procedures. We compared the simulated length (SL) from Ankyras and the labeled length (LL) provided by the vendor to the real-case postdeployment measured length. We also compared the expansion from Ankyras simulation (SE) with the real case measured expansion (ME). Results: Our analysis revealed a mean accuracy for SL across all cases of 92.05% (SD: 4.93%; range: 81.60%–99.10%), while the LL accuracy was 78.71% (SD: 12.43%; range: 62.20%–98.38%). A Pearson R² test indicated a strong correlation for SL at 0.9818, compared to 0.8625 for LL. Furthermore, the mean accuracy for expansion prediction was 86.28% (SD: 4.88%; range: 79.34%–92.46%). Conclusion: The Ankyras software shows promise as a viable tool for sizing PVST. Enhancing the accuracy of expansion predictions may further improve the precision of device specific simulation.Keywords
This publication has 17 references indexed in Scilit:
- AneuGuide™ software-assisted vs. manual measurements in sizing for pipeline embolization device: An agreement studyComputers in Biology and Medicine, 2023
- Utility of virtual stenting in treatment of cerebral aneurysms by flow diverter devicesLa radiologia medica, 2023
- Spontaneous subarachnoid haemorrhageThe Lancet, 2022
- Mechanism of Action and Biology of Flow Diverters in the Treatment of Intracranial AneurysmsNeurosurgery, 2019
- Early Results in Flow Diverter Sizing by Computational Simulation: Quantification of Size Change and Simulation Error AssessmentOperative Neurosurgery, 2018
- Unruptured intracranial aneurysms: epidemiology, natural history, management options, and familial screeningThe Lancet Neurology, 2014
- Which Spring is the Best? Comparison of Methods for Virtual StentingIEEE Transactions on Biomedical Engineering, 2014
- High Fidelity Virtual Stenting (HiFiVS) for Intracranial Aneurysm Flow Diversion: In Vitro and In SilicoAnnals of Biomedical Engineering, 2013
- Computer modeling of deployment and mechanical expansion of neurovascular flow diverter in patient-specific intracranial aneurysmsJournal of Biomechanics, 2012
- Fast virtual deployment of self-expandable stents: Method and in vitro evaluation for intracranial aneurysmal stentingMedical Image Analysis, 2012