Toward automatic C-arm positioning for standard projections in orthopedic surgery
Authors:
Open Access
- 12 June 2020
- journal article
- research article
- Published by Springer Nature in International Journal of Computer Assisted Radiology and Surgery
- Vol. 15 (7) , 1095-1105
Abstract
Purpose: Guidance and quality control in orthopedic surgery increasingly rely on intra-operative fluoroscopy using a mobile C-arm. The accurate acquisition of standardized and anatomy-specific projections is essential in this process. The corresponding iterative positioning of the C-arm is error prone and involves repeated manual acquisitions or even continuous fluoroscopy. To reduce time and radiation exposure for patients and clinical staff and to avoid errors in fracture reduction or implant placement, we aim at guiding—and in the long-run automating—this procedure.Methods: In contrast to the state of the art, we tackle this inherently ill-posed problem without requiring patient-individual prior information like preoperative computed tomography (CT) scans, without the need of registration and without requiring additional technical equipment besides the projection images themselves. We propose learning the necessary anatomical hints for efficient C-arm positioning fromin silicosimulations, leveraging masses of 3D CTs. Specifically, we propose a convolutional neural network regression model that predicts 5 degrees of freedom pose updates directly from a first X-ray image. The method is generalizable to different anatomical regions and standard projections.Results: Quantitative and qualitative validation was performed for two clinical applications involving two highly dissimilar anatomies, namely the lumbar spine and the proximal femur. Starting from one initial projection, the mean absolute pose error to the desired standard pose is iteratively reduced across different anatomy-specific standard projections. Acquisitions of both hip joints on 4 cadavers allowed for an evaluation on clinical data, demonstrating that the approach generalizes without retraining.Conclusion: Overall, the results suggest the feasibility of an efficient deep learning-based automated positioning procedure, which is trained on simulations. Our proposed 2-stage approach for C-arm positioning significantly improves accuracy on synthetic images. In addition, we demonstrated that learning based on simulations translates to acceptable performance on real X-rays.Keywords
Funding Information
- Siemens Healthineers
- Siemens Healthineers
This publication has 18 references indexed in Scilit:
- Predicting Slice-to-Volume Transformation in Presence of Arbitrary Subject Motion2017
- Pose optimization of a C-arm imaging device to reduce intraoperative radiation exposure of staff and patient during interventional procedures2017
- C-arm positioning using virtual fluoroscopy for image-guided surgery2017
- X-Ray PoseNet: 6 DoF Pose Estimation for Mobile X-Ray Devices2017
- Trends in epidemiology and patho-anatomical pattern of proximal humeral fracturesInternational Orthopaedics, 2014
- Desired-View Controlled Positioning of Angiographic C-arms2014
- The Medical Imaging Interaction Toolkit: challenges and advancesInternational Journal of Computer Assisted Radiology and Surgery, 2013
- Closed-Form Inverse Kinematics for Interventional C-Arm X-Ray Imaging With Six Degrees of Freedom: Modeling and ApplicationIEEE Transactions on Medical Imaging, 2012
- Informatics in Radiology: Use of a C-Arm Fluoroscopy Simulator to Support Training in Intraoperative RadiographyRadioGraphics, 2011
- Closed-form inverse kinematic solution for fluoroscopic C-armsAdvanced Robotics, 2007