In orthopedics, assessing the initial stability of pedicle screws is crucial for preventing failure after implantation. Traditional methods for determining stability are applicable only to post-screw insertion, complicating recovery in cases of inadequate strength. This study aimed to predict implant stability at the tapping stage prior to screw placement. Artificial bones with five distinct densities were utilized to measure tapping and screw insertion torques, facilitating a statistical evaluation of bone density and placement strength. A laser resonance frequency analysis (L-RFA) was conducted to obtain the vibration frequency spectra of the tap. Furthermore, linear approximation and support vector regression (SVR) were employed, utilizing spectral data and a 95% confidence interval to predict bone density. Significant differences in tapping and screw insertion torques were observed relative to bone density ( p R 2 of 0.843. The acquisition of bone density data at the tapping stage enables the prediction of expected screw stability. Bone density prediction using L-RFA has the potential to provide stability indicators prior to pedicle screw insertion.
Funding Information
  • Japan Society for the Promotion of Science (JP20K14684)
  • Japan Society for the Promotion of Science (JP24K15713)

This publication has 24 references indexed in Scilit: