End-to-end technical validation of multi-isocenter total body irradiation using an anthropomorphic phantom
- 24 August 2026
- journal article
- research article
- Published by Springer Nature in Radiological Physics and Technology
- Vol. 19 (4) , 1602-1616
Abstract
The objective of this study is to perform an end-to-end (E2E) technical validation of a total body irradiation (TBI) workflow using an anthropomorphic phantom, integrating simulation, treatment planning, patient-specific quality assurance, and phantom-based dosimetry to ensure accurate and reproducible dose delivery. A full E2E test was performed using the Rando Phantom and Varian Eclipse TPS. CT scans were acquired in head-first supine (HFS) and feet-first supine (FFS) orientations and registered to simulate full-body imaging. Treatment plans were optimized for both orientations, with specific gradient dose regions defined for overlap zones. ArcCHECK and OSLDs were used for pre-treatment and phantom-embedded dosimetric verification, respectively. Cone Beam CT (CBCT) assessed positional accuracy across eight treatment fractions. Dose discrepancies and alignment errors were evaluated across anatomical regions. HU comparison between the CT console and TPS revealed only minor deviations, all within ± 6 HU, supporting the consistency of HU representation in the TPS. Gamma passing rates exceeded 97% across most regions. OSLD dose differences remained within ± 2.5% for most sites, except thorax and overlap zones, which reached ± 4.8%. CBCT-based setup verification demonstrated sub-millimeter alignment deviations across fractions, indicating consistent positional accuracy. E2E validation using the Rando Phantom supports technical accuracy and consistency in the TBI workflow. Notable dose variations in overlap and heterogeneous regions highlight the need for enhanced QA in complex geometries. The findings support the technical feasibility and reproducibility of the workflow, while further patient-specific and motion-inclusive studies are required to establish broader clinical applicability.Keywords
This publication has 29 references indexed in Scilit:
- Engineering dosimetric excellence in total body irradiation: Tomotherapy‐driven protocols for precision and reproducibilityPrecision Radiation Oncology, 2026
- Comparative dosimetric analysis of base dose and gradient-based optimization in overlapping regions of extended target volumes using an anthropomorphic phantomRadiological Physics and Technology, 2025
- Implementation of superficial radiation therapy (SRT) using SRT‐100 Vision™ for non‐melanoma skin cancer in a Radiation Oncology clinicJournal of Applied Clinical Medical Physics, 2023
- Robustness of a multi-isocenter VMAT Total Body Irradiation (TBI) dose distribution against daily patient setup errorsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2022
- Adverse Effects of Total Body Irradiation: A Two-Decade, Single Institution AnalysisAdvances in Radiation Oncology, 2021
- Total Body Irradiation in Stem Cell Transplant2020
- AAPM TG 191: Clinical use of luminescent dosimeters: TLDs and OSLDsMedical Physics, 2019
- Implementing safe and robust Total Marrow Irradiation using Helical Tomotherapy – A practical guidePhysica Medica, 2019
- Tolerance limits and methodologies forIMRTmeasurement‐based verificationQA:Recommendations ofAAPMTask Group No. 218Medical Physics, 2018
- Design and implementation of an anthropomorphic quality assurance phantom for intensity-modulated radiation therapy for the Radiation Therapy Oncology GroupInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2005