The trained observer better predicts daytime ABPM diastolic blood pressure in hypertensive patients than does an automated (Omron) device
- 1 April 2006
- journal article
- research article
- Published by Wolters Kluwer Health in Blood Pressure Monitoring
- Vol. 11 (2) , 53-58
Abstract
Objectives Accurate blood pressure measurement is critical to successful clinical trials. Concerns about observer errors have led to the use of automated oscillometric devices without evidence that their performance is similar to that of trained observers. This study compares blood pressures obtained by trained observers and with an oscillometric device (Omron 705CP) to 24-h ambulatory blood pressure monitoring. Methods We performed a post-hoc analysis of 313 untreated hypertensive patients at the end of the washout phase of a Novartis hypertension trial. Patients had three seated trained observer mercury auscultatory blood pressure measurements followed by 24-h ambulatory blood pressure monitoring. The next day, the ambulatory blood pressure monitoring was removed and three seated readings were obtained with an Omron 705CP. Correlations for systolic blood pressure and diastolic blood pressure were obtained between daytime ambulatory blood pressure monitoring (0900 and 2100) and the two office methods. In addition, we investigated the degree of difference of trained observer and Omron measurements from ambulatory blood pressure monitoring. Results For systolic blood pressure, the correlation with ambulatory blood pressure monitoring of the trained observer was significantly better than with that of the Omron 705CP (0.641 vs. 0.555, P=0.01). For diastolic blood pressure values, even greater disparity between the two office method correlations with ambulatory blood pressure monitoring was observed (trained observer=0.593 vs. Omron=0.319, PPPP<0.0001 for systolic blood pressure and diastolic blood pressure). Conclusions For clinical trials using diastolic blood pressure targets, the Omron 705CP cannot replace the auscultatory blood pressure measurements of a trained observer. For systolic blood pressure, the Omron device and the trained observer had similar correlations with ambulatory blood pressure monitoring; however, both methods gave consistently higher systolic blood pressure values. Further study of oscillometric devices should be conducted before universally replacing auscultatory blood pressure determinations by trained observers in clinical trials.Keywords
This publication has 25 references indexed in Scilit:
- Recommendations for Blood Pressure Measurement in Humans and Experimental AnimalsHypertension, 2005
- Clinical decision-making in hypertension using an automated (BpTRU™) measurement deviceJournal of Human Hypertension, 2003
- Prognostic Value of Ambulatory Blood-Pressure Recordings in Patients with Treated HypertensionNew England Journal of Medicine, 2003
- The Seventh Report of the Joint National Committee on Prevention, Detection, Evaluation, and Treatment of High Blood PressureThe JNC 7 ReportJAMA, 2003
- European Society of Hypertension recommendations for conventional, ambulatory and home blood pressure measurementJournal Of Hypertension, 2003
- A Monte Carlo Evaluation of Tests for Comparing Dependent CorrelationsThe Journal of General Psychology, 2003
- 1999 World Health Organization-International Society of Hypertension Guidelines for the Management of HypertensionClinical and Experimental Hypertension, 1999
- Circadian blood pressure changes and left ventricular hypertrophy in essential hypertension.Circulation, 1990
- Sources of error in recording the blood pressure of patients with hypertension in general practice.BMJ, 1984
- Relationship Between Level of Blood Pressure Measured Casually and by Portable Recorders and Severity of Complications in Essential HypertensionCirculation, 1966