PET characteristics of a dedicated breast PET/CT scanner prototype
- 17 June 2009
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 54 (13) , 4273-4287
Abstract
A dedicated breast PET/CT system has been constructed at our institution, with the goal of having increased spatial resolution and sensitivity compared to whole-body systems. The purpose of this work is to describe the design and the performance characteristics of the PET component of this device. Average spatial resolution of a line source in warm background using maximum a posteriori (MAP) reconstruction was 2.5 mm, while the average spatial resolution of a phantom containing point sources using filtered back projection (FBP) was 3.27 mm. A sensitivity profile was computed with a point source translated across the axial field of view (FOV) and a peak sensitivity of 1.64% was measured at the center of the FOV. The average energy resolution determined on a per-crystal basis was 25%. The characteristic dead time for the front-end electronics and data acquisition (DAQ) was determined to be 145 ns and 3.6 micros, respectively. With no activity outside the FOV, a peak noise-equivalent count rate of 18.6 kcps was achieved at 318 microCi (11.766 MBq) in a cylindrical phantom of diameter 75 mm. After the effects of exposing PET detectors to x-ray flux were evaluated and ameliorated, a combined PET/CT scan was performed. The percentage standard deviations of uniformity along axial and transaxial directions were 3.7% and 2.8%, respectively. The impact of the increased reconstructed spatial resolution compared to typical whole-body PET scanners is currently being assessed in a clinical trial.Keywords
This publication has 38 references indexed in Scilit:
- Crystal identification in positron emission tomography using nonrigid registration to a Fourier-based templatePhysics in Medicine & Biology, 2008
- Dedicated Breast CT: Initial Clinical ExperienceRadiology, 2008
- The positron emission mammography/tomography breast imaging and biopsy system (PEM/PET): design, construction and phantom-based measurementsPhysics in Medicine & Biology, 2008
- Evaluation of the spatial resolution characteristics of a cone‐beam breast CT scannerMedical Physics, 2006
- High-Resolution Fluorodeoxyglucose Positron Emission Tomography with Compression ("Positron Emission Mammography") is Highly Accurate in Depicting Primary Breast CancerThe Breast Journal, 2006
- Diagnosis of Complete Response to Neoadjuvant Chemotherapy Using Diagnostic Imaging in Primary Breast Cancer PatientsThe Breast Journal, 2005
- Resolution and noise properties of MAP reconstruction for fully 3-D PETIEEE Transactions on Medical Imaging, 2000
- Comparison of 3-D maximum a posteriori and filtered backprojection algorithms for high-resolution animal imaging with microPETIEEE Transactions on Medical Imaging, 2000
- Afterglow in LSO and its possible effect on energy resolutionIEEE Transactions on Nuclear Science, 2000
- Practical cone-beam algorithmJournal of the Optical Society of America A, 1984