One Year’s Results from a Server-Based System for Performing Reject Analysis and Exposure Analysis in Computed Radiography
Authors:
- 3 November 2009
- journal article
- research article
- Published by Springer Nature in Journal of Digital Imaging
- Vol. 24 (2) , 243-255
Abstract
Rejected images represent both unnecessary radiation exposure to patients and inefficiency in the imaging operation. Rejected images are inherent to projection radiography, where patient positioning and alignment are integral components of image quality. Patient motion and artifacts unique to digital image receptor technology can result in rejected images also. We present a centralized, server-based solution for the collection, archival, and distribution of rejected image and exposure indicator data that automates the data collection process. Reject analysis program (RAP) and exposure indicator data were collected and analyzed during a 1-year period. RAP data were sorted both by reason for repetition and body part examined. Data were also stratified by clinical area for further investigation. The monthly composite reject rate for our institution fluctuated between 8% and 10%. Positioning errors were the main cause of repeated images (77.3%). Stratification of data by clinical area revealed that areas where computed radiography (CR) is seldom used suffer from higher reject rates than areas where it is used frequently. S values were log-normally distributed for examinations performed under either manual or automatic exposure control. The distributions were positively skewed and leptokurtic. S value decreases due to radiologic technology student rotations, and CR plate reader calibrations were observed. Our data demonstrate that reject analysis is still necessary and useful in the era of digital imaging. It is vital though that analysis be combined with exposure indicator analysis, as digital radiography is not self-policing in terms of exposure. When combined, the two programs are a powerful tool for quality assurance.Keywords
This publication has 15 references indexed in Scilit:
- Image Retake Analysis in Digital Radiography Using DICOM Header InformationJournal of Digital Imaging, 2008
- Digital Radiography Reject Analysis: Data Collection Methodology, Results, and Recommendations from an In-depth Investigation at Two HospitalsJournal of Digital Imaging, 2008
- Digital Repeat Analysis; Setup and OperationJournal of Digital Imaging, 2006
- Is Reject Analysis Necessary after Converting to Computed Radiography?Journal of Digital Imaging, 2002
- Comparative reject analysis in conventional film-screen and digital storage phosphor radiographyEuropean Radiology, 1999
- A comparison of image reject rates when using film, hard copy computed radiography and soft copy images on picture archiving and communication systems (PACS) workstations.British Journal of Radiology, 1999
- Potential for unnecessary patient exposure from the use of storage phosphor imaging systems1993
- Digital mobile radiographyJournal of Thoracic Imaging, 1990
- A Two-year Reject/retake Profile Analysis in Pediatric RadiologyHealth Physics, 1982
- QUALITY ASSURANCE IN SMALL DIAGNOSTIC RADIOLOGY FACILITIES.Health Physics, 1980