Evaluation of Two Total Hip Bearing Materials for Resistance to Wear Using a Hip Simulator
Open Access
- 3 June 2015
- journal article
- research article
- Published by MDPI AG in Lubricants
- Vol. 3 (2) , 459-474
Abstract
Electron beam crosslinked ultra high molecular weight polyethylene (UHMWPE) 32 mm cups with cobalt alloy femoral heads were compared with gamma-irradiation sterilized 26 mm cups and zirconia ceramic heads in a hip wear simulator. The testing was performed for a total of ten million cycles with frequent stops for cleaning and measurement of mass losses due to wear. The results showed that the ceramic on UHMWPE bearing design exhibited higher early wear than the metal on highly crosslinked samples. Once a steady state wear rate was reached, the wear rates of the two types of hip bearing systems were similar with the ceramic on UHMPWE samples continuing to show a slightly higher rate of wear than the highly crosslinked samples. The wear rates of each of the tested systems appear to be consistent with the expectations for low rates of wear in improved hip replacement systems.Keywords
Funding Information
- Masaaki Matsubara, MD, PhD
This publication has 67 references indexed in Scilit:
- Clinical Comparison of Polyethylene Wear with Zirconia or Cobalt-Chromium Femoral HeadsClinical Orthopaedics and Related Research, 2009
- Femoral Head Size and Wear of Highly Cross-linked Polyethylene at 5 to 8 YearsClinical Orthopaedics and Related Research, 2009
- Zirconia versus Co-Cr Femoral Heads in Total Hip ArthroplastyClinical Orthopaedics and Related Research, 2006
- Contact stress assessment of conventional and highly crosslinked ultra high molecular weight polyethylene acetabular liners with finite element analysis and pressure sensitive filmJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2006
- Crosslinked polyethylene compared to conventional polyethylene in total hip replacement: Pre-clinical evaluation, in-vitro testing and prospective clinical follow-up studyActa Orthopaedica, 2006
- Effect of alumina femoral head on clinical results in cemented total hip arthroplasty: old versus current aluminaJournal of Orthopaedic Science, 2005
- Clinical and simulator wear study of alumina ceramic THR to 17 years and beyondJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2003
- Polyethylene cross‐linking by two different methods reduces acetabular liner wear in a hip joint wear simulatorJournal of Orthopaedic Research, 2003
- Accelerated aging studies of UHMWPE. I. Effect of resin, processing, and radiation environment on resistance to mechanical degradationJournal of Biomedical Materials Research, 2002
- Wear of ion-implanted pure titanium against UHMWPEBiomaterials, 1989