Clinical Evaluation of a New Point-of-Care System for Chemistry Panel Testing
Authors:
- 1 September 2019
- journal article
- research article
- Published by Oxford University Press (OUP) in The Journal of Applied Laboratory Medicine
- Vol. 4 (2) , 180-192
Abstract
Background: GEM Premier ChemSTAT™ is a point-of-care (POC) system that measures Na+, K+, Ca++, Cl−, glucose, hematocrit, creatinine, blood urea nitrogen (BUN), tCO2, pH, pCO2, and lactate from a single whole blood specimen, providing rapid results in POC settings such as the emergency department (ED). Accurate measurements of creatinine in whole blood and reporting of estimated glomerular filtration rate (eGFR) can minimize adverse effects of contrast-induced nephropathy. Methods: Heparinized whole blood specimens from the ED were analyzed on the ChemSTAT by POC staff. Method comparison was performed against the cobas Integra c501 for creatinine, BUN, and tCO2, and against the GEM Premier 4000 for all other analytes. Precision was conducted with whole blood specimens assayed in triplicate over 6 days. Creatinine results from whole blood and plasma were used for eGFR, by isotope dilution mass spectrometry-traceable Modification of Diet in Renal Disease and Chronic Kidney Disease Epidemiology Collaboration equations, and eGFR concordance was assessed. Results: Creatinine, BUN, and tCO2 correlated well with plasma samples on the cobas, and all other analytes correlated well with whole blood specimens on the GEM Premier 4000 across the tested sample ranges. The regression slope was 0.951 to 1.047, along with a correlation coefficient (r) of ≥0.982 for all analytes. The pooled within-sample precision was 0% to 2.5% for all analytes. Conclusions: ChemSTAT demonstrated a strong correlation with the comparative methods and excellent precision. The system's analytical performance and continuous quality management make it suitable for use in the ED to provide rapid reliable test results, which could minimize the time to treatment and improve ED efficiency.Keywords
This publication has 15 references indexed in Scilit:
- Characteristics and Trends of Emergency Department Visits in the United States (2010–2014)The Journal of Emergency Medicine, 2019
- Early Point-of-Care Testing at Triage Reduces Care Time in Stable Adult Emergency Department PatientsThe Journal of Emergency Medicine, 2018
- Point-of-care creatinine testing for kidney function measurement prior to contrast-enhanced diagnostic imaging: evaluation of the performance of three systems for clinical utilityClinical Chemistry and Laboratory Medicine (CCLM), 2018
- Accuracy of bedside point of care testing in critical emergency department patientsThe American Journal of Emergency Medicine, 2018
- Precision treatment for cancer: Role of prognostic and predictive markersCritical Reviews in Clinical Laboratory Sciences, 2013
- Utility of point-of-care testing in ED triageThe American Journal of Emergency Medicine, 2013
- Evaluation of rapid point-of-care creatinine testing in the radiology service of a large academic medical center: Impact on clinical operations and patient dispositionClinica Chimica Acta, 2012
- Blood sample tube transporting system versus point of care technology in an emergency department; effect on time from collection to reporting? A randomised trialScandinavian Journal of Trauma, Resuscitation and Emergency Medicine, 2012
- Prevention of contrast induced nephropathy: Recommendations for the high risk patient undergoing cardiovascular proceduresCatheterization and Cardiovascular Interventions, 2006
- The Future of Emergency Care in the United States Health SystemAnnals of Emergency Medicine, 2006