Efficacy Estimation of Microbubble-Assisted Local Sonothrombolysis Using a Catheter with a Series of Miniature Transducers
Open Access
- 26 May 2021
- journal article
- research article
- Published by MDPI AG in Micromachines
- Vol. 12 (6) , 612
Abstract
Intravascular ultrasound has good prospects for clinical applications in sonothrombolysis. The catheter-based side-looking intravascular ultrasound thrombolysis (e.g., Ekosonic catheters) used in clinical studies has a high frequency (2 MHz). The lower-frequency ultrasound requires a larger-diameter transducer. In our study, we designed and manufactured a small ultrasound-based prototype catheter that can emit a lower frequency ultrasound (1.1 MHz). In order to evaluate the safety and efficacy of local low-frequency ultrasound-enhanced thrombolysis, a microbubble (MB) was introduced to augment thrombolysis effect of locally delivered low-intensity ultrasound. The results demonstrated that combination of ultrasound and MB realized higher clot lysis than urokinase-only treatment (17.0% ± 1.2% vs. 14.9% ± 2.7%) under optimal ultrasound settings of 1.1 MHz, 0.414 MPa, 4.89 W/cm2, 5% duty cycle and MB concentration of 60 μg/mL. When urokinase was added, the fibrinolysis accelerated by MB and ultrasound resulted in a further increased thrombolysis rate that was more than two times than that of urokinase alone (36.7% ± 5.5% vs. 14.9% ± 2.7%). However, a great quantity of ultrasound energy was required to achieve substantial clot lysis without MB, leading to the situation that temperature accumulated inside the clot became harmful. We suggest that MB-assisted local sonothrombolysis be considered as adjuvant therapy of thrombolytic agents.Keywords
Funding Information
- National Key Research and Development Program of China (2019YFC0120500 and 2018YFC0116201)
- the Funds of Chinese Academy of Sciences (Y201961, 20A122062ZY, and YJKYYQ20180031)
- National Natural Science Foundation of China (21927803)
This publication has 42 references indexed in Scilit:
- Laser enhanced high-intensity focused ultrasound thrombolysis: An in vitro studyThe Journal of the Acoustical Society of America, 2013
- Antithrombotic Therapy for VTE DiseaseChest, 2012
- Ultrasound-Enhanced Thrombolysis Using Definity® as a Cavitation Nucleation AgentUltrasound in Medicine & Biology, 2008
- Determinants of health‐related quality of life during the 2 years following deep vein thrombosisJournal of Thrombosis and Haemostasis, 2008
- Ultrasound-Enhanced Thrombolysis: EKOS EndoWave Infusion Catheter SystemSeminars in Interventional Radiology, 2008
- Incidence and mortality of venous thrombosis: a population‐based studyJournal of Thrombosis and Haemostasis, 2007
- Economic burden of deep-vein thrombosis, pulmonary embolism, and post-thrombotic syndromeAmerican Journal of Health-System Pharmacy, 2006
- Effect of mild hypothermia on the thrombolytic efficacy of 120 kHz ultrasound enhanced thrombolysis in an in-vitro human clot modelThrombosis Research, 2006
- Relation between quality of anticoagulant treatment and the development of the postthrombotic syndromeJournal of Thrombosis and Haemostasis, 2005
- Technology overview: sonoVueEuropean Radiology Supplements, 2004