Real-Time Passive Acoustic Mapping Using Sparse Matrix Multiplication
- 23 December 2020
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
- Vol. 68 (1) , 164-177
Abstract
Passive acoustic mapping enables the spatiotemporal monitoring of cavitation with circulating microbubbles during focused ultrasound (FUS)-mediated blood-brain barrier opening. However, the computational load for processing large data sets of cavitation maps or more complex algorithms limit the visualization in real-time for treatment monitoring and adjustment. In this study, we implemented a graphical processing unit (GPU)-accelerated sparse matrix-based beamforming and time exposure acoustics in a neuronavigation-guided ultrasound system for real-time spatiotemporal monitoring of cavitation. The system performance was tested in silico through benchmarking, in vitro using nonhuman primate (NHP) and human skull specimens, and demonstrated in vivo in NHPs. We demonstrated the stability of the cavitation map for integration times longer than 62.5 μs. A compromise between real-time displaying and cavitation map quality obtained from beamformed RF data sets with a size of 2000×128×30 (axial pixels × lateral pixels × samples) was achieved for an integration time of 1.44 μs, which required a computational time of 0.27 s (frame rate of 3.7 Hz) and could be displayed in real-time between pulses at PRF = 2 Hz. Our benchmarking tests show that the GPU sparse-matrix algorithm processed the RF data set at a computational rate of 0.03 ± 0.01 μs/pixel/sample, which enables adjusting the frame rate and the integration time as needed. The neuronavigation system with real-time implementation of cavitation mapping facilitated the localization of the cavitation activity and helped to identify distortions due to FUS phase aberration. The in vivo test of the method demonstrated the feasibility of GPU-accelerated sparse matrix computing in a close to a clinical condition, where focus distortions exemplify problems during treatment. These experimental conditions show the need for spatiotemporal monitoring of cavitation with real-time capability that enables the operator to correct or halt the sonication in case substantial aberrations are observed.Keywords
Funding Information
- National Institutes of Health (R01AG038961, R01EB009041)
This publication has 80 references indexed in Scilit:
- Low Intensity Ultrasound Promotes the Sensitivity of Rat Brain Glioma to Doxorubicin by Down-Regulating the Expressions of P-Glucoprotein and Multidrug Resistance Protein 1 In Vitro and In VivoPLOS ONE, 2013
- Real-time three-dimensional passive cavitation detection for clinical high intensity focused ultrasound systemsThe Journal of the Acoustical Society of America, 2013
- Focal Delivery of AAV2/1-transgenes Into the Rat Brain by Localized Ultrasound-induced BBB OpeningMolecular Therapy Nucleic Acids, 2013
- Blood-Brain Barrier: Real-time Feedback-controlled Focused Ultrasound Disruption by Using an Acoustic Emissions–based ControllerRadiology, 2012
- Targeted Delivery of Neural Stem Cells to the Brain Using MRI-Guided Focused Ultrasound to Disrupt the Blood-Brain BarrierPLOS ONE, 2011
- Passive cavitation mapping for localization and tracking of bubble dynamicsThe Journal of the Acoustical Society of America, 2010
- Antibodies Targeted to the Brain with Image-Guided Focused Ultrasound Reduces Amyloid-β Plaque Load in the TgCRND8 Mouse Model of Alzheimer's DiseasePLOS ONE, 2010
- Passive cavitation imaging with ultrasound arraysThe Journal of the Acoustical Society of America, 2009
- Ultrasound Enhanced Delivery of Molecular Imaging and Therapeutic Agents in Alzheimer's Disease Mouse ModelsPLOS ONE, 2008
- Noninvasive localized delivery of Herceptin to the mouse brain by MRI-guided focused ultrasound-induced blood–brain barrier disruptionProceedings of the National Academy of Sciences, 2006