Interplay effects between dose distribution quality and positioning accuracy in total marrow irradiation with volumetric modulated arc therapy
Authors:
- Pietro MancosuPMPietro Mancosu
- Piera NavarriaPNPiera Navarria
- Luca CastagnaLCLuca Castagna
- Giacomo ReggioriGRGiacomo Reggiori
- Barbara SarinaBSBarbara Sarina
- Stefano TomatisSTStefano Tomatis
- Filippo AlongiFAFilippo Alongi
- Giorgia NicoliniGNGiorgia Nicolini
- Antonella FogliataAFAntonella Fogliata
- Luca CozziLCLuca Cozzi
- Marta ScorsettiMSMarta Scorsetti
- 10 October 2013
- journal article
- Published by Wiley in Medical Physics
- Vol. 40 (11) , 111713
Abstract
Purpose:: To evaluate the dosimetric consequences of inaccurate isocenter positioning during treatment of total marrow (lymph‐node) irradiation (TMI‐TMLI) using volumetric modulated arc therapy (VMAT).Methods:: Four patients treated with TMI and TMLI were randomly selected from the internal database. Plans were optimized with VMAT technique. Planning target volume (PTV) included all the body bones; for TMLI, lymph nodes and spleen were considered into the target, too. Dose prescription to PTV was 12 Gy in six fractions, two times per day for TMI, and 2 Gy in single fraction for TMLI. Ten arcs on five isocenters (two arcs for isocenter) were used to cover the upper part of PTV (i.e., from cranium to middle femurs). For each plan, three series of random shifts with values between −3 and +3 mm and three between −5 and +5 mm were applied to the five isocenters simulating involuntary patient motion during treatment. The shifts were applied separately in the three directions: left–right (L‐R), anterior–posterior (A‐P), and cranial–caudal (C‐C). The worst case scenario with simultaneous random shifts in all directions simultaneously was considered too. Doses were recalculated for the 96 shifted plans (24 for each patient).Results:: For all shifts, differences 100 cm3. Maximum doses increased up to 15% for C‐C shifted plans. PTV covered by the 95% isodose decreased of 2%–8% revealing target underdosage with the highest values in C‐C direction.Conclusions:: The correct isocenter repositioning of TMI‐TMLI patients is fundamental, in particular in C‐C direction, in order to avoid over‐ and underdosages especially in the overlap regions. For this reason, a dedicated immobilization system was developed in the authorsˈ center to best immobilize the patient.Keywords
This publication has 18 references indexed in Scilit:
- Anatomy driven optimization strategy for total marrow irradiation with a volumetric modulated arc therapy techniqueJournal of Applied Clinical Medical Physics, 2012
- Verification of dose distribution for volumetric modulated arc therapy total marrow irradiation in a humanlike phantomMedical Physics, 2011
- Total Marrow Irradiation With RapidArc Volumetric Arc TherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2011
- Preclinical Assessment of Volumetric Modulated Arc Therapy for Total Marrow IrradiationInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2011
- Image-Guided Total-Marrow Irradiation Using Helical Tomotherapy in Patients With Multiple Myeloma and Acute Leukemia Undergoing Hematopoietic Cell TransplantationInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- Feasibility study for linac‐based intensity modulated total marrow irradiationMedical Physics, 2008
- Linac-Based Intensity Modulated Total Marrow Irradiation (IM-TMI)Technology in Cancer Research & Treatment, 2006
- Feasibility study of helical tomotherapy for total body or total marrow irradiationa)Medical Physics, 2005
- Total body irradiation before allogeneic bone marrow transplantation: is more dose better?International Journal of Radiation Oncology - Biology - Physics (IJROBP), 2001
- Allogeneic marrow transplantation in patients with chronic myeloid leukemia in the chronic phase: a randomized trial of two irradiation regimensBlood, 1991