Image quality improvements of electronic portal imaging devices by multi-level gain calibration and temperature correction
Authors:
- S HuberSHS Huber
- M MooslechnerMMM Mooslechner
- B MitterlechnerBMB Mitterlechner
- H WeichenbergerHWH Weichenberger
- M SerpaMSM Serpa
- F SedlmayerFSF Sedlmayer
- H DeutschmannHDH Deutschmann
- 2 September 2013
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 58 (18) , 6429-6446
Abstract
Amorphous silicon (aSi:H) flat panel detectors are prevalent in radiotherapy for megavoltage imaging tasks. Any clinical and dosimetrical application requires a well-defined dose response of the system to achieve meaningful results. Due to radiation damages, panels deteriorate and the linearity of pixel response to dose as well as the stability with regard to changing operating temperatures get worse with time. Using a single level gain correction can lead to an error of about 23% when irradiating a flood field image with 100 MU min(-1) on an old detector. A multi-level gain (MLG) correction is introduced, emending the nonlinearities and subpanel-related artifacts caused by insufficient radiation hardness of amplifiers in the read-out electronics. With rising temperature, offset values typically increase (up to 300 gray values) while the response at higher dose values per frame remain constant for a majority of pixels. To account for temperature-related image artifacts, two additional temperature correction methods have been developed. MLG in combination with temperature corrections can re-establish the aSi:H image quality to the performance required by reliable medical verification tools. Furthermore, the life span and recalibration intervals of these costly devices can be prolonged decisively.Keywords
This publication has 28 references indexed in Scilit:
- First Clinical Release of an Online, Adaptive, Aperture-Based Image-Guided Radiotherapy Strategy in Intensity-Modulated Radiotherapy to Correct for Inter- and Intrafractional Rotations of the ProstateInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2012
- Electron beam quality control using an amorphous silicon EPIDMedical Physics, 2009
- Developments in electronic portal imaging systemsBritish Journal of Radiology, 2006
- Portal imagingBritish Journal of Radiology, 2005
- Clinical Use of Electronic Portal ImagingSeminars in Radiation Oncology, 2005
- Multilayer Gafchromic film detectors for breast skin dose determinationin vivoPhysics in Medicine & Biology, 2002
- Clinical use of electronic portal imaging: Report of AAPM Radiation Therapy Committee Task Group 58Medical Physics, 2001
- Megavoltage imaging with a large-area, flat-panel, amorphous silicon imagerInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1996
- Demonstration of megavoltage and diagnostic x‐ray imaging with hydrogenated amorphous silicon arraysMedical Physics, 1992
- A review of electronic portal imaging devices (EPIDs)Medical Physics, 1992