Performance of MR learned pulse sequences for 3D bi‐exponential, stretched‐exponential, and mono‐exponential T2 and T1ρ mapping of knee cartilage
- 23 September 2024
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 93 (2) , 643-656
Abstract
Purpose: To compare the performance of a learned magnetization‐prepared gradient echo (L‐MPGRE) sequence against a commonly used sequence for 3D T2 and T1ρ mapping of the knee joint, the magnetization‐prepared angle‐modulated partitioned k‐space spoiled gradient echo snapshots (MAPSS), on bi‐exponential (BE), stretched‐exponential (SE), and mono‐exponential (ME) relaxation models.Methods: We used a combined differentiable and non‐differentiable optimization to learn pulse sequence structure and its parameters for 3D T2 and T1ρ mapping of the knee joint using ME, SE, and BE models. The learned pulse sequence framework was used to improve quantitative accuracy and SNR and to reduce filtering effects. We compare the measured multi‐compartment values between the two sequences (n = 8), and their repeatability (n = 4) in healthy volunteers (n = 12 total).Results: The voxel‐wise median absolute percentage difference (MAPD) between the T2 and T1ρ maps obtained with each sequence was 18.6% and 19.9%, respectively. The T2 and T1ρ repeatability tests showed a MAPD of 18.5% and 19.1% for MAPSS, and 16.8% and 15.5% for L‐MPGRE. Bland–Altman region of interest (ROI)‐wise analysis shows that bias is small, close to −1.5%, and the coefficient of variation is around 5.5% when comparing ROIs from both sequences.Conclusion: The L‐MPGRE sequences can be used as a replacement for MAPSS for T2 and T1ρ mapping in the knee cartilage with advantages, achieving similar accuracy and 15% better repeatability in only half of its scan time.Keywords
Funding Information
- National Institute of Arthritis and Musculoskeletal and Skin Diseases (P41‐EB017183, R01‐AR076328‐01A1, R01‐AR076985‐01A1, R01‐AR078308‐01A1)
- National Institute of Biomedical Imaging and Bioengineering (P41‐EB017183)
This publication has 40 references indexed in Scilit:
- Prestructural cartilage assessment using MRIJournal of Magnetic Resonance Imaging, 2016
- Extended phase graphs: Dephasing, RF pulses, and echoes ‐ pure and simpleJournal of Magnetic Resonance Imaging, 2014
- Rapid multicomponent T2 analysis of the articular cartilage of the human knee joint at 3.0TJournal of Magnetic Resonance Imaging, 2013
- Multicomponent T2 relaxation analysis in cartilageMagnetic Resonance in Medicine, 2009
- In vivo T1ρ mapping in cartilage using 3D magnetization‐prepared angle‐modulated partitioned k‐space spoiled gradient echo snapshots (3D MAPSS)Magnetic Resonance in Medicine, 2008
- Sparse MRI: The application of compressed sensing for rapid MR imagingMagnetic Resonance in Medicine, 2007
- Calculation of flip angles for echo trains with predefined amplitudes with the extended phase graph (EPG)‐algorithm: Principles and applications to hyperecho and TRAPS sequencesMagnetic Resonance in Medicine, 2003
- Phase‐Modulated binomial RF pulses for fast spectrally‐selective musculoskeletal imagingMagnetic Resonance in Medicine, 1996
- Rapid NMR imaging of dynamic processes using the FLASII techniqueMagnetic Resonance in Medicine, 1986