Real-time dose reconstruction for wedged photon beams: a generalized procedure
Open Access
- 1 September 2011
- journal article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 12 (4) , 124-138
Abstract
A practical and accurate generalized procedure to reconstruct the isocenter dose Diso for 3D conformal radiotherapy (3DCRT) has been developed for X‐ray open beams supplied by linacs of different manufacturers and equipped with aSi electronic portal imaging devices (aSi EPIDs). This paper reports an extension of the method, to be applied at the wedged X‐ray beams characterized by the wedge attenuation factor WAF.Keywords
Funding Information
- Istituto Nazionale di Fisica Nucleare applicata alla Medicina (INFN-MED)
This publication has 18 references indexed in Scilit:
- Generalized EPID calibration for in vivo transit dosimetryPhysica Medica, 2011
- Calibration of portal imaging devices for radiotherapy in-vivo dosimetryNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2010
- A literature review of electronic portal imaging for radiotherapy dosimetryRadiotherapy and Oncology, 2008
- A global calibration model for EPIDs used for transit dosimetryMedical Physics, 2007
- Application of a practical method for the isocenter pointin vivodosimetry by a transit signalPhysics in Medicine & Biology, 2007
- Routine individualised patient dosimetry using electronic portal imaging devicesRadiotherapy and Oncology, 2007
- In vivo dosimetry by an aSi‐based EPIDMedical Physics, 2006
- The use of an aSi-based EPID for routine absolute dosimetric pre-treatment verification of dynamic IMRT fieldsRadiotherapy and Oncology, 2004
- Dose-response and ghosting effects of an amorphous silicon electronic portal imaging deviceMedical Physics, 2003