Evaluation of a novel triple-channel radiochromic film analysis procedure using EBT2
Authors:
- 15 June 2012
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 57 (13) , 4353-4368
Abstract
A novel approach to read out radiochromic film was introduced recently by the manufacturer of GafChromic film. In this study, the performance of this triple-channel film dosimetry method was compared against the conventional single-red-channel film dosimetry procedure, with and without inclusion of a pre-irradiation (pre-IR) film scan, using EBT2 film and kilo- and megavoltage photon beams up to 10 Gy. When considering regions of interest averaged doses, the triple-channel method and both single-channel methods produced equivalent results. Absolute dose discrepancies between the triple-channel method, both single-channel methods and the treatment planning system calculated dose values, were no larger than 5 cGy for dose levels up to 2.2 Gy. Signal to noise in triple-channel dose images was found to be similar to signal to noise in single-channel dose images. The accuracy of resulting dose images from the triple- and single-channel methods with inclusion of pre-IR film scan was found to be similar. Results of a comparison of EBT2 data from a kilovoltage depth dose experiment to corresponding Monte Carlo depth dose data produced dose discrepancies of 9.5 ± 12 cGy and 7.6 ± 6 cGy for the single-channel method with inclusion of a pre-IR film scan and the triple-channel method, respectively. EBT2 showed to be energy sensitive at low kilovoltage energies with response differences of 11.9% and 15.6% in the red channel at 2 Gy between 50-225 kVp and 80-225 kVp photon spectra, respectively. We observed that the triple-channel method resulted in non-uniformity corrections of ±1% and consistency values of 0-3 cGy for the batches and dose levels studied. Results of this study indicate that the triple-channel radiochromic film read-out method performs at least as well as the single-channel method with inclusion of a pre-IR film scan, reduces film non-uniformity and saves time with elimination of a pre-IR film scan.Keywords
This publication has 22 references indexed in Scilit:
- High dose per fraction dosimetry of small fields with Gafchromic EBT2 filmMedical Physics, 2011
- A closer look at RapidArc® radiosurgery plans using very small fieldsPhysics in Medicine & Biology, 2011
- Characterization of image quality and image‐guidance performance of a preclinical microirradiatorMedical Physics, 2011
- A comprehensive study of the Gafchromic EBT2 radiochromic film. A comparison with EBTMedical Physics, 2010
- Technical Note: Homogeneity of Gafchromic®EBT2 filmMedical Physics, 2010
- On the characterization and uncertainty analysis of radiochromic film dosimetryMedical Physics, 2009
- Evaluation of an Epson flatbed scanner to read Gafchromic EBT films for radiation dosimetryPhysics in Medicine & Biology, 2009
- Optimizing the dynamic range extension of a radiochromic film dosimetry systemMedical Physics, 2009
- Precise radiochromic film dosimetry using a flat-bed document scannerMedical Physics, 2005
- Radiochromic film for medical radiation dosimetryMaterials Science and Engineering: R: Reports, 2003