Pre-bending a dynamic compression plate significantly alters strain distribution near the fracture plane in the mid-shaft femur
Open Access
- 5 February 2020
- journal article
- research article
- Published by SAGE Publications in Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
- Vol. 234 (5) , 478-485
Abstract
This study evaluated the effect of pre-bending dynamic compression plates on fracture site compression. Recommendations of 1 to 2 mm of pre-bend have been proposed, but there does not appear to be experimental data to confirm the optimal pre-bend magnitude. Dynamic compression plating was performed on the lateral convex surface of 18 femoral analogs to fixate a simulated mid-shaft fracture. Plates with 0 mm (flat plate), 1 mm, and 2 mm of pre-bend were evaluated for their production of compression by determining the strain magnitudes for 10 equal-sized zones across the anterior cortex at the osteotomy site using digital imaging correlation. The 0 and 1 mm plates produced significantly more compression at the near cortex (p = 0.001 and p = 0.003, respectively) than the 2 mm plate. However, the 0 and 1 mm plates also created visible diastasis at the far cortex, while the 2 mm plate exhibited compression across all zones. The strain magnitudes for the 0 mm (R2 = 0.62) and 1 mm (R2 = 0.86) plates linearly and significantly decreased from the region adjacent to the plate until a region 50%–60% across the analog diameter. In contrast, the 2 mm plate exhibited uniform strains across the osteotomy site. This study demonstrates that pre-bending a dynamic compression plate 2 mm prior to fixation on a convex lateral femur provides the most compression at the far cortex. It also produces more uniform compression across the fracture when compared to 0 and 1 mm of pre-bend.Keywords
Funding Information
- McLaren Foundation
- National Science Foundation
This publication has 19 references indexed in Scilit:
- A Comparison of the Accuracy of Three Intraoperative Techniques for Measuring Rotational Correction in Varus Derotational Osteotomies of the FemurJournal of Bone and Joint Surgery, 2014
- High Speed Fracture Fixation: Assessing Resulting Fixation Stability and Fastener Withdrawal StrengthJournal of Biomechanical Engineering, 2013
- Repeatability of digital image correlation for measurement of surface strains in composite long bonesJournal of Biomechanics, 2013
- Effect of plate working length on plate stiffness and cyclic fatigue life in a cadaveric femoral fracture gap model stabilized with a 12-hole 2.4 mm locking compression plateBMC Veterinary Research, 2013
- Experimental validation of finite element model for proximal composite femur using optical measurementsJournal of the Mechanical Behavior of Biomedical Materials, 2013
- Can DCP and LCP plates generate more compression? The effect of multiple eccentrically placed screws and their drill positioning guidesInjury, 2011
- Biomechanical Characterization of an Osteoporotic Artificial Bone Model for the Distal FemurJournal of Biomaterials Applications, 2010
- Structural properties of fourth-generation composite femurs and tibiasJournal of Biomechanics, 2008
- Internal plate fixation of fractures: short history and recent developmentsJournal of Orthopaedic Science, 2006
- Stress analysis of compression plate fixation and its effects on long bone remodelingJournal of Biomechanics, 1985