Effect of plate working length on plate stiffness and cyclic fatigue life in a cadaveric femoral fracture gap model stabilized with a 12-hole 2.4 mm locking compression plate
Open Access
- 24 June 2013
- journal article
- research article
- Published by Springer Nature in BMC Veterinary Research
- Vol. 9 (1) , 125
Abstract
Background: There are several factors that can affect the fatigue life of a bone plate, including the mechanical properties of the plate and the complexity of the fracture. The position of the screws can influence construct stiffness, plate strain and cyclic fatigue of the implants. Studies have not investigated these variables in implants utilized for long bone fracture fixation in dogs and cats. The purpose of the present study was to evaluate the effect of plate working length on construct stiffness, gap motion and resistance to cyclic fatigue of dog femora with a simulated fracture gap stabilized using a 12-hole 2.4 mm locking compression plates (LCP). Femora were plated with 12-hole 2.4 mm LCP using 2 screws per fracture segment (long working length group) or with 12-hole 2.4 mm LCP using 5 screws per fracture segment (a short working length group).Results: Construct stiffness did not differ significantly between stabilization techniques. Implant failure did not occur in any of the plated femora during cycling. Mean ± SD yield load at failure in the short plate working length group was significantly higher than in the long plate working length group.Conclusion: In a femoral fracture gap model stabilized with a 2.4 mm LCP applied in contact with the bone, plate working length had no effect on stiffness, gap motion and resistance to fatigue. The short plate working length constructs failed at higher loads; however, yield loads for both the short and long plate working length constructs were within physiologic range.Keywords
This publication has 33 references indexed in Scilit:
- Choosing a proper working length can improve the lifespan of locked platesClinical Biomechanics, 2011
- How Much Do Locked Screws Add to the Fixation of “Hybrid” Plate Constructs in Osteoporotic Bone?Journal of Orthopaedic Trauma, 2010
- The effect of the combination of locking screws and non-locking screws on the torsional properties of a locking-plate constructVeterinary and Comparative Orthopaedics and Traumatology, 2010
- Far Cortical Locking Can Reduce Stiffness of Locked Plating Constructs While Retaining Construct StrengthJournal of Bone and Joint Surgery, 2009
- A review of locking compression plate biomechanics and their advantages as internal fixators in fracture healingClinical Biomechanics, 2007
- Fatigue Analysis of Plates Used for Fracture Stabilization in Small Dogs and CatsVeterinary Surgery, 2006
- Biomechanical testing of the LCP – how can stability in locked internal fixators be controlled?Injury, 2003
- The Strength of Plate Fixation in Relation to the Number and Spacing of Bone ScrewsJournal of Orthopaedic Trauma, 1996
- The Effects of Screw Removal on Bone Strain in an Idealized Plated Bone ModelVeterinary Surgery, 1988
- Cancellous Bone Strength at the Proximal Human TibiaEngineering in Medicine, 1984