Positioning of a plane-parallel ionization chamber in clinical electron beams and the impact on perturbation factors
Authors:
- K ZinkKZK Zink
- J WulffJWJ Wulff
Open Access
- 1 April 2009
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 54 (8) , 2421-2435
Abstract
Current dosimetry protocols recommend the use of plane-parallel ionization chambers for the dosimetry of clinical electron beams. The necessary perturbation corrections p(wall) and p(cav) are assumed to be unity, independent of the depth of measurement and the energy of the primary electrons. To verify these assumptions detailed Monte Carlo studies of a Roos chamber in clinical electron beams with energies in the range of 6-21 MeV are performed at different depths in water and analyzed in terms of Spencer-Attix cavity theory. Separate simulations for the perturbation corrections p(wall) and p(cav) indicate quite different properties of both correction factors with depth. Dose as well as fluence calculations show a nearly depth-independent wall correction factor for a shift of the Roos chamber Deltaz = -0.017 cm toward the focus. This value is in good agreement with the positioning recommendation given in all dosimetry protocols. Regarding the fluence perturbation p(cav) the simulation of the electron fluence inside the air cavity in comparison to water unambiguously reveals an in-scattering of low energy electrons, despite the fact, that the cavity is 'well guarded'. For depths beyond the reference depth z(ref) this effect is superimposed by an increased loss of primary electrons from the beam resulting in p(cav) > 1. This effect is largest for low electron energies but present for all electron energies involved in this study. Based on the different depth dependences of p(wall) and p(cav) it is possible to choose a chamber shift Deltaz in a way to minimize the depth dependence of the overall perturbation factor p. For the Roos chamber this shift is Deltaz = -0.04 cm independent of electron energy.Keywords
This publication has 22 references indexed in Scilit:
- Monte Carlo calculations of correction factors for plane‐parallel ionization chambers in clinical electron dosimetryMedical Physics, 2008
- On the effective point of measurement in megavoltage photon beamsMedical Physics, 2006
- Wall correction factors, Pwall, for parallel-plate ionization chambersMedical Physics, 2006
- Investigation of variance reduction techniques for Monte Carlo photon dose calculation using XVMCPhysics in Medicine & Biology, 2000
- Accurate condensed history Monte Carlo simulation of electron transport. II. Application to ion chamber response simulationsMedical Physics, 2000
- Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 versionMedical Physics, 2000
- AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon and electron beamsMedical Physics, 1999
- R50 as a beam quality specifier for selecting stopping‐power ratios and reference depths for electron dosimetryMedical Physics, 1996
- Calculation of stopping‐power ratios using realistic clinical electron beamsMedical Physics, 1995
- Einfluß der Vielfachstreuung von Elektronen auf die Ionisation in gasgefüllten HohlräumenRadiation and Environmental Biophysics, 1968