Slice‐specific B1+ shimming improves the repeatability of multishot DWI at 7 T
Open Access
- 1 August 2024
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
Abstract
Purpose: Compared with lower field strengths, DWI at 7 T faces the combined challenges of increased distortion and blurring due to B0 inhomogeneity, and increased signal dropouts due to B1+ inhomogeneity. This study addresses the B1+ limitations using slice‐specific static parallel transmission (pTx) in a multi‐shot, readout‐segmented EPI diffusion imaging sequence.Methods: DWI was performed in 7 healthy subjects using MRI at 7 T and readout‐segmented EPI. Data were acquired with non‐pTx circular‐polarized (CP) pulses (CP‐DWI) and static pTx pulses (pTx‐DWI) using slice‐specific B1+ shim coefficients. Each volunteer underwent two scan sessions on the same day, with two runs of each sequence in the first session and one run in the second. The sequences were evaluated by assessing image quality, flip‐angle homogeneity, and intrasession and intersession repeatability in ADC estimates.Results: pTx‐DWI significantly reduced signal voids compared with CP‐DWI, particularly in inferior brain regions. The use of pTx also improved RF uniformity and symmetry across the brain. These effects translated into improved intrasession and intersession repeatability for pTx‐DWI. Additionally, re‐optimizing the pTx pulse between repeat scans did not have a negative effect on ADC repeatability.Conclusion: The study demonstrates that pTx provides a reproducible image‐quality increase in multishot DWI at 7 T. The benefits of pTx also extend to quantitative ADC estimation with regard to the improvement in intrasession and intersession repeatability. Overall, the combination of multishot imaging and pTx can support the development of reliable, high‐resolution DWI for clinical studies at 7 T.Keywords
This publication has 40 references indexed in Scilit:
- Left–Right Intensity Asymmetries Vary Depending on Scanner Model for FLAIR and T1 Weighted MRI ImagesJournal of Magnetic Resonance Imaging, 2022
- RF coils: A practical guide for nonphysicistsJournal of Magnetic Resonance Imaging, 2018
- Abnormal asymmetry of white matter integrity in schizophrenia revealed by voxelwise diffusion tensor imagingHuman Brain Mapping, 2011
- PROPELLER EPI: An MRI technique suitable for diffusion tensor imaging at high field strength with reduced geometric distortionsMagnetic Resonance in Medicine, 2005
- High‐resolution diffusion‐weighted imaging with interleaved variable‐density spiral acquisitionsJournal of Magnetic Resonance Imaging, 2005
- Nonlinear phase correction for navigated diffusion imagingMagnetic Resonance in Medicine, 2003
- Transmit SENSEMagnetic Resonance in Medicine, 2002
- Generalized autocalibrating partially parallel acquisitions (GRAPPA)Magnetic Resonance in Medicine, 2002
- Diffusion‐weighted interleaved echo‐planar imaging with a pair of orthogonal navigator echoesMagnetic Resonance in Medicine, 1996
- Acute human stroke studied by whole brain echo planar diffusion‐weighted magnetic resonance imagingAnnals of Neurology, 1995