Abstract
Goldmann applanation tonometry (GAT) is the internationally accepted standard for intra-ocular pressure (IOP) measurement, which is important for the diagnosis of glaucoma. The technique does not consider the effect of the natural variation in the corneal thickness, curvature and material properties. As these parameters affect the structural resistance of the cornea, their variation is expected to lead to inaccuracies in IOP determination. Numerical Analysis based on the finite element method has been used to simulate the loading conditions experienced in GAT and hence assess the effect of variation in corneal parameters on GAT IOP measurements. The analysis is highly nonlinear and considers the hyper-elastic J-shaped stress–strain properties of corneal tissue observed in laboratory tests. The results reveal a clear association between both the corneal thickness and material properties, and the measured IOP. Corneal curvature has a considerably lower effect. Similar trends have been found from analysis of clinical data involving 532 patients referred to the Glaucoma Unit at Moorfields Hospital, and from earlier mathematical analyses. Nonlinear modelling is shown to trace the behaviour of the cornea under both IOP and tonometric pressure, and to be able to provide additional, and potentially useful, information on the distribution of stress, strain, contact pressure and gap closure.Keywords
This publication has 28 references indexed in Scilit:
- Numerical modelling of corneal biomechanical behaviourComputer Methods in Biomechanics and Biomedical Engineering, 2007
- The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometryBritish Journal of Ophthalmology, 2005
- Comparative study of corneal strip extensometry and inflation testsJournal of The Royal Society Interface, 2005
- Influence of corneal biomechanical properties on intraocular pressure measurementJournal of Cataract & Refractive Surgery, 2005
- Application of structural analysis to the mechanical behaviour of the corneaJournal of The Royal Society Interface, 2004
- Effect of corneal parameters on measurements using the pulsatile ocular blood flow tonograph and Goldmann applanation tonometerBritish Journal of Ophthalmology, 2004
- Central corneal thickness and intraocular pressure in a Mongolian population,Ophthalmology, 1998
- Comparison of mechanical properties of keratoconus and normal corneasExperimental Eye Research, 1982
- The Application of Newton’s Method to the Problem of Elastic StabilityJournal of Applied Mechanics, 1972
- Über ApplanationstonometrieOphthalmologica, 1957