Common-Mode Interference in Biopotential Amplifiers: Modeling, Analysis, and Design Strategies for Various Recording Setups
- 23 October 2025
- journal article
- review article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Circuits and Systems
- Vol. 20 (1) , 3-14
Abstract
This review provides a comprehensive overview of techniques for mitigating common-mode interference (CMI) in biopotential analog front-ends (AFEs). The mechanisms of CMI generation in various biopotential measurement scenarios, including neurostimulation and two-electrode ECG, are modeled electrically. The impact of CMI on signal quality is analyzed from both small-signal and large-signal views, highlighting the scenario-dependent nature of the CMI issue. Techniques for improving the common-mode rejection ratio (CMRR) are introduced to suppress CMI in small-signal conditions. The concept of total CMRR (TCMRR), which incorporates the effect of asymmetric contact impedance, is reviewed, and corresponding design strategies for maximizing TCMRR are analyzed. In the large-signal view, CMI-induced distortion and approaches for enhancing tolerance CMI are discussed, addressing both sub-supply and over-supply CMI scenarios. By analyzing mitigation techniques across different measurement contexts, this review offers practical design insights to guide future biopotential AFE designers in selecting the most appropriate solutions.Keywords
Funding Information
- Korea government (RS-2025-02214408)
This publication has 33 references indexed in Scilit:
- Design of Low Power and Low Noise Instrumentation Amplifier for Biopotential Acquisition: A ReviewIEEE Access, 2025
- A 7.5GΩ-Input-Impedance 109dB-Total-CMRR and 80VPP-CMI-Tolerance Instrumentation Amplifier for Dry-Contact Two-Electrode ECG Monitoring2023
- A 2-Electrode ECG Amplifier with 0.5% Nominal Gain Shift and 0.13% THD in a 530mVpp Input Common-Mode Range2020
- 26.1 A 4.5mm2 Multimodal Biosensing SoC for PPG, ECG, BIOZ and GSR Acquisition in Consumer Wearable Devices2020
- 34.6 EEG Dust: A BCC-Based Wireless Concurrent Recording/Transmitting Concentric Electrode2020
- A 0.6V 3.8μW ECG/bio-impedance monitoring IC for disposable health patch in 40nm CMOS2018
- Capacitive Biopotential Measurement for Electrophysiological Signal Acquisition: A ReviewIEEE Sensors Journal, 2016
- An 8-channel fully differential analog front-end for neural recording2012
- A fully integrated, untrimmed CMOS instrumentation amplifier with submicrovolt offsetIEEE Journal of Solid-State Circuits, 1999
- Simple, fast method for patient body capacitance and power-line electric interference measurementMedical & Biological Engineering & Computing, 1991