A full optically operated magnetometer array: An experimental study
- 1 November 2012
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 83 (11) , 113106
Abstract
We show the operation of an optically pumped magnetometer array in a 50 μT magnetic field. The various components for a fully optical and non-magnetic detector unit were constructed and evaluated, from which a prototype unit was assembled with fiber coupled electronics. In this unit the magnetometers were operated using the intensity modulated method and heated with an off-resonant laser. Calculations on the temperature distribution were used to design the magnetometer array. Different magnetometers in such a detector unit were characterized and showed identical performance. Without applying noise reduction schemes, the obtained magnetic field resolution is a factor 2.5 above the shot noise level down to frequencies of about 7 Hz.Keywords
This publication has 11 references indexed in Scilit:
- Characteristics and performance of an intensity-modulated optically pumped magnetometer in comparison to the classical M_x magnetometerOptics Express, 2012
- Magnetoencephalography with a chip-scale atomic magnetometerBiomedical Optics Express, 2012
- Low temperature SQUID magnetometer systems for geophysical exploration with transient electromagneticsSuperconductor Science and Technology, 2011
- Microfabricated atomic vapor cell arrays for magnetic field measurementsReview of Scientific Instruments, 2011
- Noise reduction in optically pumped magnetometer assembliesApplied Physics B, 2010
- Microfabricated atomic clocks and magnetometersJournal of Optics A: Pure and Applied Optics, 2006
- Chip-scale atomic magnetometerApplied Physics Letters, 2004
- Microfabricated alkali atom vapor cellsApplied Physics Letters, 2004
- A subfemtotesla multichannel atomic magnetometerNature, 2003
- Principles of Operation of the Rubidium Vapor MagnetometerApplied Optics, 1962