Manual vs. automatic capture management in implantable cardioverter defibrillators and cardiac resynchronization therapy defibrillators
- 14 March 2010
- journal article
- clinical trial
- Published by Oxford University Press (OUP) in EP Europace
- Vol. 12 (6) , 811-816
Abstract
The Secura™ ICD and Consulta™ CRT-D are the first defibrillators to have automatic right atrial (RA), right ventricular (RV), and left ventricular (LV) capture management (CM). Complete CM was evaluated in an implantable cardioverter defibrillator (ICD) population. Two prospective clinical studies were conducted in 28 centres in Europe and Israel. Automatic CM data were compared with manual threshold measurements, the CM applicability was determined, and adjustments to pacing outputs were analysed. In total, 160 patients [age 64.6 ± 10.4 years, 77% male, 80 ICD and 80 cardiac resynchronization therapy defibrillator (CRT-D)] were included. The differences between automatic and manual measurements were ≤0.25 V in 97% (RA CM) and 96% (RV CM) and were all within the safety margin. Fully automatic CM measurements were available within 1 week prior to the 3-month visit in 90% (RA), 99% (RV), and 97% (LV) of the patients. Results indicated increased output (threshold >2.5 V) due to raised RA threshold in seven (4.4%), high RV threshold in nine (5.6%), and high LV threshold in three patients (3.8%). All high threshold detections and all automatic modulations of pacing output were adjudicated appropriate. Complete CM adjusts pacing output appropriately, permitting a reduction in office visits while it may maximize device longevity. The study was registered at ClinicalTrials.gov identifiers: NCT00526227 and NCT00526162.Keywords
This publication has 17 references indexed in Scilit:
- Fluctuation of left ventricular thresholds and required safety margin for left ventricular pacing with cardiac resynchronization therapyEP Europace, 2009
- Automatic Management of Left Ventricular Stimulation: Hints for Technologic ImprovementPacing and Clinical Electrophysiology, 2009
- Atrial Threshold Variability: Implications for Automatic Atrial Stimulation AlgorithmsPacing and Clinical Electrophysiology, 2007
- Evolution of pacing for bradycardia: AutocaptureEuropean Heart Journal Supplements, 2007
- Automated Left Ventricular Capture ManagementPacing and Clinical Electrophysiology, 2007
- Ventricular Pacing Threshold Variations in the YoungPacing and Clinical Electrophysiology, 2007
- ACC/AHA/ESC 2006 Guidelines for Management of Patients With Ventricular Arrhythmias and the Prevention of Sudden Cardiac DeathJACC, 2006
- Role of ventricular autocapture function in increasing longevity of DDDR pacemakers: a prospective studyEP Europace, 2006
- Automatic measurement of atrial pacing thresholds in dual-chamber pacemakers: Clinical experience with atrial capture managementHeart Rhythm, 2005
- Capture management efficacy in children and young adults with endocardial and unipolar epicardial systemsEP Europace, 2004