Predictive value of short-term migration in determining long-term stable fixation in cemented and cementless total knee arthroplasties
Authors:
- 1 July 2019
- journal article
- conference paper
- Published by British Editorial Society of Bone & Joint Surgery in The Bone & Joint Journal
- Vol. 101-B (7_Supple_C) , 55-60
Abstract
Aims: Early implant migration measured with radiostereometric analysis (RSA) has been proposed as a useful predictor of long-term fixation of tibial components in total knee arthroplasty. Evaluation of actual long-term fixation is of interest for cemented components, as well as for cementless fixation, which may offer long-term advantages once osseointegration has occurred. The objective of this study was to compare the long-term migration with one- and two-year migration to evaluate the predictive ability of short-term migration data and to compare migration and inducible displacement between cemented and cementless (porous metal monoblock) components at least ten years postoperatively. Patients and Methods: Patients who had participated in RSA migration studies with two-year follow-up were recruited to return for a long-term follow-up, at least ten years from surgery. Two cemented tibial designs from two manufacturers and one porous metal monoblock cementless tibial design were studied. At the long-term follow-up, patients had supine RSA examinations to determine migration and loaded examinations (single leg stance) to determine inducible displacement. In total, 79 patients (54 female) returned, with mean time since surgery of 12 years (10 to 14). There were 58 cemented and 21 cementless tibial components. Results: Migration at one year and two years was significantly correlated with long-term migration (p < 0.001). Median migration at the long-term follow-up was 0.6 mm (maximum total point motion; interquartile range (IQR) 0.4 to 0.9) for the cemented group and 0.6 mm (IQR 0.3 to 1.1) for the cementless group with no difference between groups (p = 0.99). Inducible displacement was significantly lower for the cementless components (p < 0.001). Conclusion: Long-term migration was strongly correlated with two-year migration. Although long-term migration was not different for cemented or cementless tibial components, inducible displacement at the long-term visit was significantly lower for these cementless components, suggesting superior fixation. These findings support the predictive value of short-term migration in determining long-term fixation. Cite this article: Bone Joint J 2019;101-B(7 Supple C):55–60Keywords
This publication has 28 references indexed in Scilit:
- Early migration of tibial components is associated with late revisionActa Orthopaedica, 2012
- Differences in long-term fixation between mobile-bearing and fixed-bearing knee prostheses at ten to 12 years’ follow-upThe Journal of Bone and Joint Surgery. British volume, 2012
- The beneficial effect of hydroxyapatite lastsActa Orthopaedica, 2012
- Continued stabilization of trabecular metal tibial monoblock total knee arthroplasty components at 5 years—measured with radiostereometric analysisActa Orthopaedica, 2011
- Uncemented HA-coated Implant is the Optimum Fixation for TKA in the Young PatientClinical Orthopaedics and Related Research, 2006
- Guidelines for standardization of radiostereometry (RSA) of implantsActa Orthopaedica, 2005
- Inducible displacements of cemented tibial components during weight‐bearing and knee extension. Observations during dynamic radiostereometry related to joint positions and 2 years history of migration in 16 TKRJournal of Orthopaedic Research, 2001
- Editorial Comments: Introducing New Technology: A Stepwise AlgorithmSpine, 2000
- Hydroxyapatite coating versus cemented fixation of the tibial component in total knee arthroplastyThe Journal of Arthroplasty, 1999
- The stability of three different cementless tibial components: A randomized radiostereometric study in 45 knee arthroplasty patientsActa Orthopaedica, 1995