Characterization of photodetector temporal response for neutron time-of-flight (nToF) diagnostics at the National Ignition Facility
- 1 October 2018
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 89 (10) , 10I135
Abstract
The temporal response of a microchannel plate photomultiplier tube used in the suite of neutron time of flight (nToF) diagnostics at the National Ignition Facility has been characterized to reduce uncertainty in, and understanding of, shot parameters obtained from nTOF data. A short pulse laser, neutral density glass filters, and electrical attenuators were used to gather statistically significant samples of photodetector impulse response functions (IRF) in rapid succession. Individual components have been absolutely calibrated to minimize systematic uncertainties. The zeroth (collected charge), first (transit time), and second central moments (transit time spread) of the IRF were calculated as either the bias voltage or the amount of light incident on the detector was varied. Timing reference was provided by a monitor photodiode viewing a pickoff of the incident laser pulse. The primary sources of uncertainty are jitter in the monitor photodiode and the statistical variation across our measurement period. The spreads in the first moment, with respect to the timing photodiode, and the square root of the second central moment were found to be less than 50 ps and 150 ps, respectively.Keywords
Funding Information
- U.S. Department of Energy (DE-AC52-07NA27344)
This publication has 4 references indexed in Scilit:
- A fused silica Cherenkov radiator for high precision time-of-flight measurement of DT γ and neutron spectra (invited)Review of Scientific Instruments, 2018
- Impulse responses of visible phototubes used in National Ignition Facility neutron time of flight diagnosticsReview of Scientific Instruments, 2016
- Analysis of the neutron time-of-flight spectra from inertial confinement fusion experimentsJournal of Applied Physics, 2015
- A New Neutron Time-of-Flight Detector to Measure the MeV Neutron Spectrum at the National Ignition FacilityPlasma and Fusion Research, 2014