Assessment of an updated double‐vasopressor automated system using Nexfin ™ for the maintenance of haemodynamic stability to improve peri‐operative outcome during spinal anaesthesia for caesarean section
Authors:
- 28 January 2015
- journal article
- Published by Wiley in Anaesthesia
- Vol. 70 (6) , 691-698
Abstract
Summary: Hypotension occurs commonly during spinal anaesthesia for caesarean section, associated with maternal and fetal adverse effects. We developed a double‐vasopressor automated system with a two‐step algorithm and continuous non‐invasive haemodynamic monitoring using the Nexfin ™ device. The system delivered 25 μg phenylephrine every 30 s when systolic blood pressure was between 90% and 100% of baseline, or 2 mg ephedrine at this blood pressure range and heart rate < 60 beats.min −1 ; and 50 μg phenylephrine or 4 mg ephedrine when systolic blood pressure was < 90% of baseline with the same heart rate criterion. Fifty‐seven women received standardised spinal anaesthesia. Twenty‐seven (47.4%) had at least one reading of hypotension defined as systolic blood pressure < 80% baseline. Systolic blood pressure was within 20% of the baseline in a mean ( SD ) of 79.8 (20.9)% of measurements. Fifty‐three (93.0%) women required phenylephrine before delivery while 10 (17.5%) required ephedrine. Six women (10.5%) experienced nausea and three (5.3%) vomited. The system was able to achieve a low incidence of maternal hypotension with good maternal and fetal outcomes.Keywords
Funding Information
- SingHealth Foundation Grant (SHF/CTG047/2012)
This publication has 12 references indexed in Scilit:
- Closed‐loop double‐vasopressor automated system vs manual bolus vasopressor to treat hypotension during spinal anaesthesia for caesarean section: a randomised controlled trialAnaesthesia, 2013
- Closed‐loop double‐vasopressor automated system to treat hypotension during spinal anaesthesia for caesarean section: a preliminary studyAnaesthesia, 2012
- Noninvasive Continuous Arterial Blood Pressure Monitoring with Nexfin®Anesthesiology, 2012
- A comparison of the Nexfin ® and transcardiopulmonary thermodilution to estimate cardiac output during coronary artery surgeryAnaesthesia, 2012
- Prevention of maternal hypotension after regional anaesthesia for caesarean sectionCurrent Opinion in Anaesthesiology, 2010
- Low-dose spinal anaesthesia for Caesarean section to prevent spinal-induced hypotensionCurrent Opinion in Anaesthesiology, 2008
- Closed‐loop feedback computer‐controlled infusion of phenylephrine for maintaining blood pressure during spinal anaesthesia for caesarean section: a preliminary descriptive study*Anaesthesia, 2007
- Vasopressors in obstetrics: what should we be using?Current Opinion in Anaesthesiology, 2006
- Prevention of Hypotension during Spinal Anesthesia for Cesarean DeliveryAnesthesiology, 2005
- Measuring the predictive performance of computer-controlled infusion pumpsJournal of Pharmacokinetics and Biopharmaceutics, 1992