Abstract
This paper proposes a chopper-stabilized three-opamp IA with an active high-pass filter (HPF) for biological signal acquisition. An equivalent resistance amplification technique is employed to construct a second-order high-pass filtering network, which significantly reduces the area on the chip occupied by resistors and capacitors and improves the level of circuit integration. In addition, an improved chopper modulation architecture is adopted to realize low-frequency noise and offset suppression, thus enabling high-precision signal acquisition. The proposed IA is fabricated in a standard 180 nm CMOS process. Under a 3.3 V supply voltage, the power consumption is only 1.281 mW. Leveraging the inherent high-frequency gain roll-off characteristic of the operational amplifier, the designed IA achieves a -3 dB bandwidth of 0.05 Hz to 20 kHz, which fully covers the frequency range of mainstream bioelectronic signals such as electrocardiogram (ECG) and electroencephalogram (EEG). Measurements results show that IA achieves a noise of 28.524 nV/ √ Hz @100 Hz, a common-mode rejection ratio (CMRR) as high as 100 dB, and an input impedance of 13.3 GΩ @100 Hz. Meanwhile, it provides a highly accurate adjustable gain with a maximum value of 1000. The input-referred noise (IRN) of this IA is 0.182 μVrms within the 0.5–100 Hz and 0.215 μVrms within the 0.05–150 Hz, achieving competitive noise efficiency factor (3.20) and power efficiency factor (33.79).
Funding Information
  • Shenzhen Science and Technology Program (202206193000001)

This publication has 22 references indexed in Scilit: