Image quality and radiation dose comparison of prospectively triggered low-dose CCTA: 128-slice dual-source high-pitch spiral versus 64-slice single-source sequential acquisition
- 10 July 2011
- journal article
- research article
- Published by Springer Nature in The International Journal of Cardiovascular Imaging
- Vol. 28 (5) , 1217-1225
Abstract
Currently 64-multislice computed tomography (MSCT) scanners are the most widely used devices allowing low radiation dose coronary CT angiography (CCTA) with prospective ECG triggering. Latest 128-slice dual-source CT (DSCT) scanners offer prospective high-pitch spiral acquisition covering the heart during one single beat. We compared radiation dose and image quality from prospective 64-MSCT versus high-pitch spiral 128-slice DSCT scanning, as such data is lacking. CCTA of 50 consecutive patients undergoing 128-DSCT (2 × 64 × 0.6 mm collimation, 0.28 s rotation time, 3.4 pitch, 100–120 kV tube voltage and 320 mAs tube current–time product) were compared to CCTA of 50 heart rate (HR) and BMI matched patients undergoing 64-MSCT (64 × 0.625 mm collimation, 0.35 s rotation time, 100–120 kV tube voltage and 400–650 mA tube current). Image quality was rated on a 4-point scale by two independent cardiac imaging physicians (1 = excellent to 4 = non-diagnostic). Of 710 coronary segments assessed on 128-DSCT, 216 (30.4%) achieved an image quality score 1 excellent, 400 (56.3%) score 2, 76 (10.7%) score 3 and 18 (2.6%) score 4 (non-diagnostic). Of 737 coronary segments evaluated on 64-MSCT 271 (36.8%) had an image quality score of 1, 327 (44.4%) 2, 110 (14.9%) score 3, and 29 (3.9%) segments score 4. Average image quality score for both scanners was similar (P = 0.641). The mean heart rate during scanning was 58.7 ± 5.6 bpm on 128-DSCT and 59.0 ± 5.6 bpm on 64-MSCT, respectively. Mean effective radiation dose was 1.0 ± 0.2 mSv for 128-DSCT and 1.7 ± 0.6 mSv for 64-MSCT (P < 0.001). 128-DSCT with high-pitch spiral mode allows CCTA acquisition with reduced radiation dose at maintained image quality compared to 64-MSCT.Keywords
This publication has 25 references indexed in Scilit:
- High-pitch dual-source CT coronary angiography: systolic data acquisition at high heart ratesEuropean Radiology, 2010
- Low Dose High-Pitch Spiral Acquisition 128-Slice Dual-Source Computed Tomography for the Evaluation of Coronary Artery Bypass Graft PatencyInvestigative Radiology, 2010
- Diagnostic accuracy of high-pitch dual-source CT for the assessment of coronary stenoses: first experienceEuropean Radiology, 2009
- Low-Dose Coronary CT Angiography With Prospective ECG Triggering: Validation of a Contrast Material Protocol Adapted to Body Mass IndexAmerican Journal of Roentgenology, 2009
- First head-to-head comparison of effective radiation dose from low-dose 64-slice CT with prospective ECG-triggering versus invasive coronary angiographyHeart, 2009
- Body physique and heart rate variability determine the occurrence of stair-step artefacts in 64-slice CT coronary angiography with prospective ECG-triggeringEuropean Radiology, 2009
- Low-dose CT Coronary Angiography Using Prospective ECG-TriggeringAcademic Radiology, 2009
- Low-Dose Computed Tomography Coronary Angiography With Prospective Triggering: A Promise for the FutureJACC, 2008
- Radiation Dose to Patients From Cardiac Diagnostic ImagingCirculation, 2007
- A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc Committee for Grading of Coronary Artery Disease, Council on Cardiovascular Surgery, American Heart AssociationCirculation, 1975