Anterior cervical interbody constructs: effect of a repetitive compressive force on the endplate
Open Access
- 14 October 2011
- journal article
- research article
- Published by Wiley in Journal of Orthopaedic Research
- Vol. 30 (4) , 587-592
Abstract
Graft subsidence following anterior cervical reconstruction can result in the loss of sagittal balance and recurring foraminal stenosis. This study examined the implant–endplate interface using a cyclic fatigue loading protocol in an attempt to model the subsidence seen in vivo. The superior endplate from 30 cervical vertebrae (C3 to T1) were harvested and biomechanically tested in axial compression with one of three implants: Fibular allograft; titanium mesh cage packed with cancellous chips; and trabecular metal. Each construct was cyclically loaded from 50 to 250 N for 10,000 cycles. Nondestructive cyclic loading of the cervical endplate–implant construct resulted in a stiffer construct independent of the type of the interbody implant tested. The trabecular metal construct demonstrated significantly more axial stability and significantly less subsidence in comparison to the titanium mesh construct. Although the allograft construct resulted in more subsidence than the trabecular metal construct, the difference was not significant and no difference was found when comparing axial stability. For all constructs, the majority of the subsidence during the cyclic testing occurred during the first 500 cycles and was followed by a more gradual settling in the remaining 9,500 cycles. © 2011 Orthopaedic Research Society. © 2011 Orthopaedic Research Society Published by Wiley Periodicals, Inc. J Orthop Res 30:587–592, 2012Keywords
This publication has 44 references indexed in Scilit:
- Correlation of Radiostereometric Measured Cervical Range of Motion With Clinical Radiographic Findings After Anterior Cervical Discectomy and FusionSpine, 2009
- Anterior Cervical Fusion With Tantalum ImplantSpine, 2008
- Correlation of cervical endplate strength with CT measured subchondral bone densityEuropean Spine Journal, 2007
- A Comparison of Fatigue Failure Responses of Old Versus Middle-Aged Lumbar Motion Segments in Simulated Flexed LiftingSpine, 2007
- Loss of Cervical Endplate Integrity Following Minimal Surface PreparationSpine, 2007
- Factors affecting sagittal malalignment due to cage subsidence in standalone cage assisted anterior cervical fusionEuropean Spine Journal, 2007
- Mechanical damage to the intervertebral disc annulus fibrosus subjected to tensile loadingJournal of Biomechanics, 2005
- Maintenance of Interbody Space in One- and Two-Level Anterior Cervical Interbody FusionClinical Orthopaedics and Related Research, 2005
- Correlation between sagittal plane changes and adjacent segment degeneration following lumbar spine fusionEuropean Spine Journal, 2001
- Subsidence Resulting From Simulated Postoperative Neck MovementsSpine, 2000