Determination of 10 perfluorinated carboxylic acid compounds in water by gas chromatography-mass spectrometry coupled with negative chemical ionization

Authors:

Abstract
A method was developed for the simultaneous determination of 10 perfluorinated carboxylic acid compounds in water by gas chromatography-mass spectrometry coupled with negative chemical ionization ( GC-NCI-MS). Perfluorinated carboxylic acid compounds were derivatizedby trifluoro-N-methyl-N-(trimethylsilyl) acetamide ( MSTFA) as the trimethylsilyl derivatization reagent. The water sample was purified and enriched through a weak anion exchange solid phase extraction column and analyzed via GC-NCI-MS. The sample pretreatment, derivation and instrument conditions were optimized. The results showed that the linearity of the 10 perfluorinated carboxylic acid compounds was good in the range of 0. 1-10 mg/L with correlation coefficients of 0. 995 6-0.999 3. The limits of detection (LODs) and limits of quantification (LOQs) were 0. 5-1. 5 mu g/L and 1.5-4.5 mu g/L, respectively. The spiked recoveries of the blank samples ranged from 70.2% to 112.6% with the relative standard deviations (RSDs) between 2. 1% and 14.5% (n=6). The method is simple, sensitive, accurate and precise, and can be used to detect the 10 perfluorinated carboxylic acid compounds in water.