Abstract
In vivo preclinical X-ray micro-computed tomography (μCT) imaging is widely used to obtain three-dimensional anatomical information of small animals. However, the potential for radiation exposure to influence experimental outcomes necessitates accurate dose estimation. The aim of this study was to develop a robust and reproducible method for estimating and evaluating absorbed doses in small animals undergoing preclinical in vivo μCT imaging. Absorbed doses were measured at four separate positions within a dedicated polymethyl methacrylate phantom using MCP-N (LiF, Cu, P) thermoluminescence dosemeters (TLDs) and a novel type of optically stimulated luminescence dosemeter (OSLD) made from household salt (NaCl). The findings were benchmarked against the vendor’s absorbed dose estimates, revealing a discrepancy between the measured and provided values. The results for the OSLDs followed a similar trend to the TLDs, though significant statistical differences were found between the luminescent dosemeters and the vendor values. In this study, a method for measuring the absorbed doses from μCT systems was presented. The results indicate a need for thorough dose measurements prior to performing longitudinal imaging studies.Keywords
Funding Information
- Swedish Research Council (2018-02438)
This publication has 33 references indexed in Scilit:
- Dosimetry concepts for scanner quality assurance and tissue dose assessment in micro-CTMedical Physics, 2012
- Current Status of Developments and Applications of Micro-CTAnnual Review of Biomedical Engineering, 2011
- Phantom and cadaver measurements of dose and dose distribution in micro-CT of the chest in miceActa Radiologica, 2011
- Evaluation of the radiation dose in micro‐CT with optimization of the scan protocolContrast Media & Molecular Imaging, 2010
- SpekCalc: a program to calculate photon spectra from tungsten anode x-ray tubesPhysics in Medicine & Biology, 2009
- Household salt as a retrospective dosemeter using optically stimulated luminescenceRadiation and Environmental Biophysics, 2008
- TLD assessment of mouse dosimetry during microCT imagingMedical Physics, 2008
- Small Animal Absorbed Radiation Dose from Serial Micro-Computed Tomography ImagingMolecular Imaging and Biology, 2007
- Monte Carlo simulations of dose from microCT imaging procedures in a realistic mouse phantomMedical Physics, 2005
- Fundamental image quality limits for microcomputed tomography in small animalsMedical Physics, 2003