Assessment of small volume ionization chambers as reference dosimeters in high-energy photon beams
Authors:
- L de CarlanLCL de Carlan
- F DelaunayFDF Delaunay
- M DonoisMDM Donois
- P FournierPFP Fournier
- A OstrowskyAOA Ostrowsky
- A VouillaumeAVA Vouillaume
- J M BordyJBJ M Bordy
- 10 August 2011
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 56 (17) , 5637-5650
Abstract
LNE-LNHB is involved in a European project aiming at establishing absorbed dose-to-water standards for photon-radiation fields down to 2 × 2 cm². This requires the calibration of reference ionization chambers of small volume. Twenty-four ionization chambers of eight different types with volume ranging from 0.007 to 0.057 cm³ were tested in a ⁶⁰Co beam. For each chamber, two major characteristics were investigated: (1) the stability of the measured current as a function of the irradiation time under continuous irradiation. At LNE-LNHB, the variation of the current should be less than ±0.1% in comparison with its first value (over a 16 h irradiation time); (2) the variation of the ionization current with the applied polarizing voltage and polarity. Leakage currents were also measured. Results show that (1) every tested PTW (31015, 31016 and 31014) and Exradin A1SL chambers demonstrate a satisfying stability under irradiation. Other types of chambers have a stability complying with the stability criterion for some or none of them. (2) IBA CC01, IBA CC04 and Exradin A1SL show a proper response as a function of applied voltage for both polarities. PTW, Exradin A14SL and Exradin A16 do not. Only three types of chambers were deemed suitable as reference chambers according to LNE-LNHB requirements and specifications from McEwen (2010 Med. Phys. 37 2179-93): Exradin A1SL chambers (3/3), IBA CC04 (2/3) and IBA CC01 (1/3). The Exradin A1SL type with an applied polarizing voltage of 150 V was chosen as an LNE-LNHB reference chamber type in 2 × 2 cm² radiation fields.Keywords
This publication has 13 references indexed in Scilit:
- Monte Carlo calculations of , the beam quality conversion factorMedical Physics, 2010
- Ion recombination for ionization chamber dosimetry in a helical tomotherapy unitMedical Physics, 2010
- Measurement of ionization chamber absorbed dose factors in megavoltage photon beamsMedical Physics, 2010
- Traceability of absorbed dose in photon radiation therapy: from primary standard to patient treatmentAccreditation and Quality Assurance, 2009
- The determination of beam quality correction factors: Monte Carlo simulations and measurementsPhysics in Medicine & Biology, 2009
- Polarity effect in commercial ionization chambers used in photon beams with small fieldsRadiological Physics and Technology, 2009
- Response to high-energy photons of PTW31014 PinPoint ion chamber with a central aluminum electrodeMedical Physics, 2008
- Comparison of ionization chambers of various volumes for IMRT absolute dose verificationMedical Physics, 2003
- Saturation current and collection efficiency for ionization chambers in pulsed beamsMedical Physics, 2000
- AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon and electron beamsMedical Physics, 1999