Pneumatic Tactile Sensor Probe with Variable Sensitivity for Palpation during Catheterization

Authors:

Abstract
Palpation technologies in teleoperated catheterization are essential for remotely assessing tissue characteristics such as stiffness and elasticity, which are critical for identifying pathways in cardiovascular or pulmonary system, as well as to detect abnormalities like tumors. In this study, we present the development of a tactile sensor probe designed to support catheterization procedures. The sensor utilizes a pneumatic sensing mechanism, with its deformable tip fabricated from polydimethylsiloxane (PDMS) elastomer for biocompatibility and flexibility. Tip deformation is transmitted to a remotely located pressure sensor through a dual-fluid system using incompressible water and compressible air to relay force information. This configuration enables spatial separation between the sensing tip and the detection electronics, enhancing safety and compatibility in medical environments. Sensitivity of the system was tuned by adjusting the air gap height, and experimental results closely matched theoretical models. A reliable correlation was observed between sensor output, tip deformation, and applied normal force, confirming the sensor's capability for quantitative tactile feedback. These results demonstrate the potential of the proposed sensor for enhancing palpation in catheter-based, minimally invasive procedures.