Integrity of the stem‐cement interface in THA: Effects of stem surface finish and cement porosity
Authors:
- 3 April 2008
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research Part B: Applied Biomaterials
- Vol. 87B (1) , 77-82
Abstract
No consensus exists for the optimal surface finish on cemented total hip prosthesis stems. The purpose of this study was to determine the effects of stem finish and interfacial cement porosity on the integrity of the stem–cement interface. Simulated stems made of Co‐Cr, having polished or matte surfaces, at room temperature or heated to 37°C, were cemented into Sawbones simulated femurs. Push out testing of the stem–cement interface was performed immediately after cement polymerization and after two aging periods in 37°C saline or 37°C air, and the extent of interfacial porosity at the stem–cement interface was determined. Polished stems exhibited an average 60% greater interfacial strength than that of matte stems initially and up to 240% after aging treatments. Cement porosity at the stem–cement interface and incomplete cement interdigitation into surface asperities on matte stems likely allowed saline penetration into the stem–cement interface during wet aging, resulting in a rapid decrease of shear strength. Stem preheating to 37°C virtually eliminated interfacial pores and resulted in greater shear strengths regardless of surface finish. Polished stem surfaces with stem preheating provided the best interfacial shear strength and sealing ability against saline penetration into the stem–cement interface and could result in increased longevity of stem fixation. © 2008 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 2008Keywords
This publication has 38 references indexed in Scilit:
- The effects of fluid penetration and interfacial porosity on the fixation of cemented femoral componentsThe Journal of Bone and Joint Surgery. British volume, 2005
- Migration of polyethylene debris along well‐fixed cemented implantsJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2003
- Stem surface roughness alters creep induced subsidence and ‘taper-lock’ in a cemented femoral hip prosthesisJournal of Biomechanics, 2001
- Influence of Cement Technique on the Interface Strength of Femoral ComponentsClinical Orthopaedics and Related Research, 2000
- Failed innovation in total hip replacement: Diagnosis and proposals for a cureActa Orthopaedica, 1993
- Extensive porosity at the cement–femoral prosthesis interface: A preliminary studyJournal of Biomedical Materials Research, 1993
- A three-dimensional non-linear finite element study of the effect of cement-prosthesis debonding in cemented femoral total hip componentsJournal of Biomechanics, 1991
- The effect of surface preparation on metal/bone cement interfacial strengthJournal of Biomedical Materials Research, 1987
- Intramedullary fixation of artificial hip joints with bone cement‐precoated implants. I. Interfacial strengthsJournal of Biomedical Materials Research, 1982
- Factors affecting surgical alloy/bone cement interface adhesionJournal of Biomedical Materials Research, 1980