Failure of SCS MR-Conditional Modes Due to High Impedance: A Review of Literature and Case Series
Open Access
- 21 November 2020
- journal article
- research article
- Published by Springer Nature in Pain and Therapy
- Vol. 10 (1) , 729-737
Abstract
Introduction Magnetic resonance imaging (MRI) conditional modes are a novel feature for certain Food and Drug Administration (FDA)-approved spinal cord stimulation (SCS) devices. However, there is a paucity of literature around the limitation of MRI-conditional modes (“MRI safe”), specifically in clinical scenarios where urgent MRIs may be needed. One such limitation is load impedance, referring to the circuit’s resistance to the current being generated by the system. High impedance can limit the MRI-conditional mode capability, presenting potential harm to a patient undergoing an MRI or make an MRI unable to be completed. Methods Three cases were identified, and informed consent was obtained. All information was obtained via retrospective chart review. Results In this case series of three patients where MRI-conditional SCS systems were unable to be placed in “MRI safe” settings, preventing timely MRI study completion in the setting of high impedance, all three were required to undergo alternative imaging including CT scans, and two patients ultimately had the system explanted and one chose to be re-implanted after completion of scans. Conclusion This case series highlights the need for further investigation of impedance in SCS systems and potential limitations for future MRI usage. The review of literature of impedance in SCS shows both device- and physiologic-related etiologies for changes in impedance that warrant consideration by the implanting physician.Keywords
This publication has 29 references indexed in Scilit:
- Improving Patient Experience with Spinal Cord Stimulation: Implications of Position-Related Changes in NeurostimulationNeuromodulation: Technology at the Neural Interface, 2014
- Epidural Lipomatosis as a Cause for High Impedance Values During a Spinal Cord Stimulator TrialPM&R, 2013
- Current-controlled deep brain stimulation reduces in vivo voltage fluctuations observed during voltage-controlled stimulationClinical Neurophysiology, 2010
- Retrospective Review of 707 Cases of Spinal Cord Stimulation: Indications and ComplicationsPain Practice, 2010
- Magnetic Resonance Imaging in Patients with Spinal Neurostimulation SystemsAnesthesiology, 2007
- Factors Affecting Impedance of Percutaneous Leads in Spinal Cord StimulationNeuromodulation: Technology at the Neural Interface, 2006
- Modern Pacemaker and Implantable Cardioverter/Defibrillator Systems Can Be Magnetic Resonance Imaging SafeCirculation, 2004
- Effects of Posture on Stimulation Parameters in Spinal Cord StimulationNeuromodulation: Technology at the Neural Interface, 1998
- Postural Changes in Spinal Cord Stimulation Perceptual ThresholdsNeuromodulation: Technology at the Neural Interface, 1998
- Stimulation Characteristics, Complications, and Efficacy of Spinal Cord Stimulation Systems in Patients with Refractory Angina: A Prospective Feasibility StudyPacing and Clinical Electrophysiology, 1994