Evaluation of a MR‐quadrupole imaging coil for spinal interventions in a vertical 1.0 T MRI
Open Access
- 27 December 2011
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 68 (2) , 600-605
Abstract
The in vivo pain treatment was successfully performed with the patient in a prone position. The PD-weighted TSE with echo time = 10 ms rendered contrast-to-noise-ratio values of 27 ± 10 for needle/fat, 1.6 ± 5 for needle/muscle, and 4 ± 4.7 for needle/nerve tissue. The mean diameter of the needle artifact was 1.2 ± 0.2 mm. In the T1-weighted gradient echo, the needle's artifact diameter was 6 ± 2 mm; the needle's contrast-to-noise ratio relative to muscle tissue was 4 ± 2, 7.6 ± 1.5 for needle/fat, and 5 ± 1 for needle/nerve tissue. With the PD-weighted TSE (echo time = 10 ms) and the T1-weighted gradient echo, the needle was imaged reliably throughout the intervention. The butterfly surface coil is feasible for the guidance of spinal interventions in a prone patient. Magn Reson Med, 2012.Keywords
This publication has 23 references indexed in Scilit:
- MR-guided discography using an open 1 Tesla MRI systemEuropean Radiology, 2010
- Development of a signal‐inducing bone cement for magnetic resonance imagingJournal of Magnetic Resonance Imaging, 2010
- Image-guided spinal injection procedures in open high-field MRI with vertical field orientation: feasibility and technical featuresEuropean Radiology, 2009
- Freehand Real-Time MRI-Guided Lumbar Spinal Injection Procedures at 1.5 T: Feasibility, Accuracy, and SafetyAmerican Journal of Roentgenology, 2009
- Diagnostic and Interventional MRI of the Sacroiliac Joints Using a 1.5-T Open-Bore Magnet: A One-Stop-Shopping ApproachAmerican Journal of Roentgenology, 2008
- Magnetic resonance–guided percutaneous laser disk decompression for lumbar disk herniation—relationship between clinical results and location of needle tipSurgical Neurology, 2007
- Percutaneous MR‐guided discography in a low‐field system using optical instrument tracking: A feasibility studyJournal of Magnetic Resonance Imaging, 2003
- MRI-guided periradicular nerve root infiltration therapy in low-field (0.23-T) MRI system using optical instrument trackingEuropean Radiology, 2002
- An integrated head-holder/Coil for intraoperative MRI in open neurosurgeryJournal of Magnetic Resonance Imaging, 2000
- MRI-guided treatment of degenerative spinal diseasesMinimally Invasive Therapy & Allied Technologies, 1999